Reconfigurable Voltage/Current Readout Instrumentation Amplifier for Cardiovascular Health Monitoring
Authors:
- 1 May 2018
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 1-4
Abstract
A single common instrumentation amplifier (IA) architecture to measure voltage and current input signals is presented and applied to electrocardiogram (ECG) and photoplethysmogram (PPG) sensing. The reconfigurable IA switches between the voltage and current sensing modes in timely multiplexing manner faster than ECG and PPG bandwidth. A design to reduce noise is applied in the current sensing mode. The designed IA is implemented using CMOS 180-nm process technology. The current consumption, voltage noise, and current noise are measured as 45 μA, 2.7 μVrms, and 384 pArms, respectively. In addition, the measured PPG signals are compared with/without the proposed noise reduction scheme.Keywords
This publication has 3 references indexed in Scilit:
- Reconfigurable Multiparameter Biosignal Acquisition SoC for Low Power Wearable PlatformSensors, 2016
- A Multi(bio)sensor Acquisition System With Integrated Processor, Power Management, $8 \times 8$ LED Drivers, and Simultaneously Synchronized ECG, BIO-Z, GSR, and Two PPG ReadoutsIEEE Journal of Solid-State Circuits, 2016
- A Survey on Wearable Sensor-Based Systems for Health Monitoring and PrognosisIEEE Transactions on Systems, Man and Cybernetics, Part C (Applications and Reviews), 2009