AI software detection of large vessel occlusion stroke on CT angiography: a real-world prospective diagnostic test accuracy study
- 27 January 2022
- journal article
- research article
- Published by BMJ in Journal of NeuroInterventional Surgery
- Vol. 15 (1) , 52-56
Abstract
Background Artificial intelligence (AI) software is increasingly applied in stroke diagnostics. However, the actual performance of AI tools for identifying large vessel occlusion (LVO) stroke in real time in a real-world setting has not been fully studied. Objective To determine the accuracy of AI software in a real-world, three-tiered multihospital stroke network. Methods All consecutive head and neck CT angiography (CTA) scans performed during stroke codes and run through an AI software engine (Viz LVO) between May 2019 and October 2020 were prospectively collected. CTA readings by radiologists served as the clinical reference standard test and Viz LVO output served as the index test. Accuracy metrics were calculated. Results Of a total of 1822 CTAs performed, 190 occlusions were identified; 142 of which were internal carotid artery terminus (ICA-T), middle cerebral artery M1, or M2 locations. Accuracy metrics were analyzed for two different groups: ICA-T and M1 ±M2. For the ICA-T/M1 versus the ICA-T/M1/M2 group, sensitivity was 93.8% vs 74.6%, specificity was 91.1% vs 91.1%, negative predictive value was 99.7% vs 97.6%, accuracy was 91.2% vs 89.8%, and area under the curve was 0.95 vs 0.86, respectively. Detection rates for ICA-T, M1, and M2 occlusions were 100%, 93%, and 49%, respectively. As expected, the algorithm offered better detection rates for proximal occlusions than for mid/distal M2 occlusions (58% vs 28%, p=0.03). Conclusions These accuracy metrics support Viz LVO as a useful adjunct tool in stroke diagnostics. Fast and accurate diagnosis with high negative predictive value mitigates missing potentially salvageable patients.Keywords
This publication has 21 references indexed in Scilit:
- Fast Automatic Detection of Large Vessel Occlusions on CT AngiographyStroke, 2019
- Guidelines for the Early Management of Patients With Acute Ischemic Stroke: 2019 Update to the 2018 Guidelines for the Early Management of Acute Ischemic StrokeStroke, 2019
- Machine Learning–Enabled Automated Determination of Acute Ischemic Core From Computed Tomography AngiographyStroke, 2019
- Artificial intelligence to diagnose ischemic stroke and identify large vessel occlusions: a systematic reviewJournal of NeuroInterventional Surgery, 2019
- Automated Detection of Intracranial Large Vessel Occlusions on Computed Tomography AngiographyStroke, 2019
- Efficacy of endovascular thrombectomy in patients with M2 segment middle cerebral artery occlusions: meta-analysis of data from the HERMES CollaborationJournal of NeuroInterventional Surgery, 2019
- Abstract WMP16: Artificial Intelligence Detection of Cerebrovascular Large Vessel Occlusion - Nine Month, 650 Patient Evaluation of the Diagnostic Accuracy and Performance of the Viz.ai LVO AlgorithmStroke, 2019
- STARD 2015 guidelines for reporting diagnostic accuracy studies: explanation and elaborationBMJ Open, 2016
- Optimization modeling to maximize population access to comprehensive stroke centersNeurology, 2015
- A systematic review classifies sources of bias and variation in diagnostic test accuracy studiesJournal of Clinical Epidemiology, 2013