HIGH PRECISION MICROCALORIMETRY: APPARATUS, PROCEDURES, AND BIOCHEMICAL APPLICATIONS

Authors:
  • DSD.K. Steckler
    D.K. Steckler
  • RGR. N. Goldberg
    R. N. Goldberg
  • YTY.B. Tewari
    Y.B. Tewari
  • TBT.J. Buckley
    T.J. Buckley

Abstract
Apparatus and procedures used for high-precision microcalorimetric measurements are described. The calorimeter is of the heat-conduction type and utilizes semi-conductor thermoelectric modules. The bicompartmental reaction vessel is made of high-density polyethylene and holds about 0.5 mL of solution in each compartment. Imprecision of heat measurement is 0.2 percent when measuring 300 mJ of heat produced by a rapid chemical reaction. Three microcalorimeters are operated simultaneously using a microcomputer and a data acquisition system. Thermochemical and kinetic applications are described. The acquisition of data from an isoperibol solution calorimeter is also described.