Development of a compact linear accelerator to generate ultrahigh dose rate high‐energy X‐rays for FLASH radiotherapy applications
Authors:
- 19 January 2023
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 50 (3) , 1680-1698
Abstract
Purpose: In recent years, the FLASH effect, in which ultrahigh dose rate (UHDR) radiotherapy (RT) can significantly reduce toxicity to normal tissue while maintaining antitumor efficacy, has been verified in many studies and even applied in human clinical cases. This work evaluates whether a room‐temperature radio‐frequency (RF) linear accelerator (linac) system can produce UHDR high‐energy X‐rays exceeding a dose rate of 40 Gy/s at a clinical source‐surface distance (SSD), exploring the possibility of a compact and economical clinical FLASH RT machine suitable for most hospital treatmentrooms.Methods: A 1.65 m long S‐band backward‐traveling‐wave (BTW) electron linac was developed to generate high‐current electron beams, supplied by a commercial klystron‐based power source. A tungsten‐copper electron‐to‐photon conversion target for UHDR X‐rays was designed and optimized with Monte Carlo (MC) simulations using Geant4 and thermal finite element analysis (FEA) simulations using ANSYS. EBT3 and EBT‐XD radiochromic films, which were calibrated with a clinical machine Varian VitalBeam, were used for absolute dose measurements. A PTW ionization chamber detector was used to measure the relative total dose and a plane‐parallel ionization chamber detector was used to measure the relative normalized dose of each pulse.Results: The BTW linac generated 300‐mA‐pulse‐current 11 MeV electron beams with 29 kW mean beam power, and the conversion target could sustain this high beam power within a maximum irradiation duration of 0.75 s. The mean energy of the produced X‐rays was 1.66 MeV in the MC simulation. The measured flat‐filter‐free (FFF) maximum mean dose rate of the room‐temperature linac exceeded 80 Gy/s at an SSD of 50 cm and 45 Gy/s at an SSD of 67.9 cm, both at a 2.1 cm depth of the water phantom. The FFF radiation fields at 50 cm and 67.9 cm SSD at a 2.1 cm depth of the water phantom showed Gaussian‐like distributions with 14.3 and 20 cm full‐width at half‐maximum (FWHM) values, respectively.Conclusion: This work demonstrated the feasibility of UHDR X‐rays produced by a room‐temperature RF linac, and explored the further optimization of system stability. It shows that a simple and compact UHDR X‐ray solution can be facilitated for both FLASH‐RT scientific research and clinical applications.Keywords
Funding Information
- Ministry of Science and Technology of the People's Republic of China
This publication has 21 references indexed in Scilit:
- Treatment of a first patient with FLASH-radiotherapyRadiotherapy and Oncology, 2019
- PHASER: A platform for clinical translation of FLASH cancer radiotherapyRadiotherapy and Oncology, 2019
- Biological Benefits of Ultra-high Dose Rate FLASH Radiotherapy: Sleeping Beauty AwokenClinical Oncology, 2019
- The Advantage of FLASH Radiotherapy Confirmed in Mini-pig and Cat-cancer PatientsClinical Cancer Research, 2019
- X-rays can trigger the FLASH effect: Ultra-high dose-rate synchrotron light source prevents normal brain injury after whole brain irradiation in miceRadiotherapy and Oncology, 2018
- Line focus x-ray tubes—a new concept to produce high brilliance x-raysPhysics in Medicine & Biology, 2017
- (P003) Delivery of Ultra-Rapid Flash Radiation Therapy and Demonstration of Normal Tissue Sparing After Abdominal Irradiation of MiceInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2017
- Ultrahigh dose-rate FLASH irradiation increases the differential response between normal and tumor tissue in miceScience Translational Medicine, 2014
- Intensity-modulated radiation therapy for head and neck cancer: Systematic review and meta-analysisRadiotherapy and Oncology, 2014
- Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verificationJournal of Applied Clinical Medical Physics, 2013