Commissioning of a double-focused micro multileaf collimator (μMLC)
Authors:
Open Access
- 1 March 2010
- journal article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 11 (2) , 81-91
Abstract
Double‐focused μMLCs are able to create fields with steeper dose gradients at the field edges and are, therefore, an advancement in delivering stereotactic treatments. A double‐focused μMLC has been installed at a Siemens Primus linear accelerator (linac) as a first research installation in Europe. The basic dosimetric parameters, such as leakage, output factors, depth‐dose curves and penumbra, have been measured in 6 and 15 MV‐mode by use of radiochromic films (GafChromic EBT), ionization chambers and our solid water QA‐phantom (Easy Cube). The leakage between the leaves is minimal and lower than that of most commercially available MLCs. Therefore, the field size of the linac can be kept constant while the leaves of the μMLC are creating different aperture shapes. Percentage depth doses (PDDs) generated by the double‐focused μMLC are equal to depth‐dose curves of the original linac. That means the μMLC affects only the off‐axis ratio (OAR). Based on the fact that the μMLC is double‐focused and the source‐to‐collimator distance is larger, the penumbra is sharper than that for fields defined by the original linac MLC. The mechanical and dosimetric investigations show the benefit of the double‐focused μMLC attached to a Siemens Primus linear accelerator.Keywords
Funding Information
- Deutsche Forschungsgemeinschaft
- SCHM (1070/24)
This publication has 16 references indexed in Scilit:
- Radiation transmission, leakage and beam penumbra measurements of a micro‐multileaf collimator using GafChromic EBT filmJournal of Applied Clinical Medical Physics, 2008
- Comparison of advanced irradiation techniques with photons for benign intracranial tumoursRadiotherapy and Oncology, 2006
- Absorption spectra variations of EBT radiochromic film from radiation exposurePhysics in Medicine & Biology, 2005
- A dosimetric comparison of two multileaf collimator designsMedical Physics, 2002
- What is the optimum leaf width of a multileaf collimator?Medical Physics, 2000
- Dosimetric characteristics of a double-focused miniature multileaf collimatorMedical Physics, 1999
- Intensity-modulated arc therapy with dynamic multileaf collimation: an alternative to tomotherapyPhysics in Medicine & Biology, 1995
- Computer systems and mechanical tools for stereotactically guided conformation therapy with linear acceleratorsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1992
- Algorithm for dosimetry of multiarc linear-accelerator stereotactic radiosurgeryMedical Physics, 1991
- Measurements of dose distributions in small beams of 6 MV X-raysPhysics in Medicine & Biology, 1987