High-resolution short range ion detectors based on 4H-SiC films

Authors:

Abstract
The energy resolution of SiC detectors has been studied in application to the spectrometry of α particles with 5.1–5.5 MeV energies. The Schottky barrier structure of the detector was based on a CVD-grown epitaxial n-4H-SiC film with a thickness of 26 μm and an uncompensated donor concentration of (1–2)×1015 cm−3. An energy resolution of 0.5% achieved for the first time with SiC detectors allows fine structure of the α particle spectrum to be revealed. The average energy of the electron-hole pair formation in 4H-SiC is estimated at 7.71 eV.

This publication has 4 references indexed in Scilit: