Commissioning of a Quantulus 1220™ Liquid Scintillation Beta Spectrometer for Measuring 14C and 3H at Natural Abundance Levels
- 1 January 1997
- journal article
- Published by Cambridge University Press (CUP) in Radiocarbon
- Vol. 40 (1) , 201-209
Abstract
In 1994, the Gliwice Radiocarbon Laboratory began operating a liquid scintillation spectrometry system, consisting of a Quantulus 1220™ spectrometer and two vacuum rigs for benzene production. This paper describes the procedures used for the benzene synthesis from samples containing < 1 g of carbon and in the range 1 to 10 g of carbon. We also present the Quantulus calibration procedures used in the Gliwice Radiocarbon Laboratory and NERC Radiocarbon Laboratory, and compare the calibration parameters.Keywords
This publication has 6 references indexed in Scilit:
- Newly Designed 0.8-ML Teflon® Vial for Microvolume Radiocarbon DatingRadiocarbon, 1995
- Liquid Scintillation Counter Characterization, Optimization and Benzene Purity CorrectionRadiocarbon, 1992
- Four Decades of Progress in 14C Dating by Liquid Scintillation Counting and Spectrometry1992
- The Radiocarbon Laboratory at the University of BarcelonaRadiocarbon, 1991
- Multiparameter liquid scintillation spectrometryNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 1984