Axillary Temperature, as Recorded by the iThermonitor WT701, Well Represents Core Temperature in Adults Having Noncardiac Surgery
Authors:
- 1 March 2018
- journal article
- research article
- Published by Wolters Kluwer Health in Anesthesia & Analgesia
- Vol. 126 (3) , 833-838
Abstract
BACKGROUND: Core temperature can be accurately measured from the esophagus or nasopharynx during general anesthesia, but neither site is suitable for neuraxial anesthesia. We therefore determined the precision and accuracy of a novel wireless axillary thermometer, the iThermonitor, to determine its suitability for use during neuraxial anesthesia and in other patients who are not intubated. METHODS: We enrolled 80 adults having upper abdominal surgery with endotracheal intubation. Intraoperative core temperature was measured in distal esophagus and was estimated at the axilla with a wireless iThermonitor WT701 (Raiing Medical, Boston MA) at 5-minute intervals. Pairs of axillary and reference distal esophageal temperatures were compared and summarized using linear regression and repeated-measured Bland–Altman methods. We a priori determined that the iThermonitor would have clinically acceptable accuracy if most estimates were within ±0.5°C of the esophageal reference, and suitable precision if the limits of agreement were within ±0.5°C. RESULTS: There were 3339 sets of paired temperatures. Axillary and esophageal temperatures were similar, with a mean difference (esophageal minus axillary) of only 0.14°C ± 0.26°C (standard deviation). The Bland–Altman 95% limits of agreement were reasonably narrow, with the estimated upper limit at 0.66°C and the lower limit at −0.38°C, thus ±0.52°C, indicating good agreement across the range of mean temperatures from 34.9°C to 38.1°C. The absolute difference was within 0.5°C in 91% of the measurements (95% confidence interval, 88%–93%). CONCLUSIONS: Axillary temperature, as recorded by the iThermonitor WT701, well represents core temperature in adults having noncardiac surgery and thus appears suitable for clinical use.Keywords
This publication has 23 references indexed in Scilit:
- Estimation of the concordance correlation coefficient for repeated measures using SAS and RComputer Methods and Programs in Biomedicine, 2013
- Noninvasive Temperature Monitoring in Postanesthesia Care UnitsAnesthesiology, 2009
- The Concordance Correlation Coefficient for Repeated Measures Estimated by Variance ComponentsJournal of Biopharmaceutical Statistics, 2009
- Urinary bladder and oesophageal temperatures correlate better in patients with high rather than low urinary flow rates during non-cardiac surgery2008
- Temperature Monitoring and Perioperative ThermoregulationAnesthesiology, 2008
- Agreement Between Methods of Measurement with Multiple Observations Per IndividualJournal of Biopharmaceutical Statistics, 2007
- Malignant hyperthermiaOrphanet Journal of Rare Diseases, 2007
- Perioperative Heat BalanceAnesthesiology, 2000
- Isoflurane Anesthesia and Circadian Temperature Cycles in HumansAnesthesiology, 1991
- Does Urinary Catheter Temperature Reflect Core Temperature during Cardiac Surgery?Anesthesiology, 1988