Pressure flow characteristics of the valve in spacer devices.
Authors:
Open Access
- 1 September 1989
- journal article
- research article
- Published by BMJ in Archives of Disease in Childhood
- Vol. 64 (9) , 1305-1307
Abstract
The valve system in the mouth piece of two spacer devices was analysed. Pressures required to open and close the non-rebreathing valve were very low (less than 0.1 kPa). Inspiratory flow requirements were within physiological limits for infants' normal tidal breathing. Expiratory flow requirements varied significantly, and the flow required to prevent rebreathing from the chamber may exceed the physiological flow limits for normal tidal breathing.Keywords