Compressed sensing based cone‐beam computed tomography reconstruction with a first‐order methoda)
- 31 August 2010
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 37 (9) , 5113-5125
Abstract
Purpose: This article considers the problem of reconstructing cone‐beam computed tomography (CBCT) images from a set of undersampled and potentially noisy projection measurements.Methods: The authors cast the reconstruction as a compressed sensing problem based on norm minimization constrained by statistically weighted least‐squares of CBCT projection data. For accurate modeling, the noise characteristics of the CBCT projection data are used to determine the relative importance of each projection measurement. To solve the compressed sensing problem, the authors employ a method minimizing total‐variation norm, satisfying a prespecified level of measurement consistency using a first‐order method developed by Nesterov.Results: The method converges fast to the optimal solution without excessive memory requirement, thanks to the method of iterative forward and back‐projections. The performance of the proposed algorithm is demonstrated through a series of digital and experimental phantom studies. It is found a that high quality CBCT image can be reconstructed from undersampled and potentially noisy projection data by using the proposed method. Both sparse sampling and decreasing x‐ray tube current (i.e., noisy projection data) lead to the reduction of radiation dose in CBCT imaging.Conclusions: It is demonstrated that compressed sensing outperforms the traditional algorithm when dealing with sparse, and potentially noisy, CBCT projection views.Keywords
Funding Information
- National Science Foundation (0854492)
- National Institutes of Health (1R01 CA104205)
This publication has 46 references indexed in Scilit:
- Retrospective IMRT Dose Reconstruction Based on Cone-Beam CT and MLC Log-FileInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- Computed Tomography — An Increasing Source of Radiation ExposureNew England Journal of Medicine, 2007
- The management of imaging dose during image‐guided radiotherapy: Report of the AAPM Task Group 75Medical Physics, 2007
- Dose delivered from Varian's CBCT to patients receiving IMRT for prostate cancerPhysics in Medicine & Biology, 2007
- Evaluation of on-board kV cone beam CT (CBCT)-based dose calculationPhysics in Medicine & Biology, 2007
- Overview of image-guided radiation therapyMedical Dosimetry, 2006
- Patient dose from kilovoltage cone beam computed tomography imaging in radiation therapyMedical Physics, 2006
- Radiation dose reduction in four‐dimensional computed tomographyMedical Physics, 2005
- Flat-panel cone-beam computed tomography for image-guided radiation therapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2002
- Practical cone-beam algorithmJournal of the Optical Society of America A, 1984