Blood pressure in the neonate: Three non‐invasive oscillometric pressure monitors compared with invasively measured blood pressure
Authors:
- 1 February 2005
- journal article
- clinical trial
- Published by Wiley in Acta Paediatrica
- Vol. 94 (2) , 191-196
Abstract
Aim: To compare blood pressure in neonates obtained from three different oscillometer blood pressure monitors (Dinamap™, Criticare™, Hewlett‐Packard™) with arterial blood pressure (Hewlett‐Packard™ invasive unit). Study design: A total of 32 randomized, non‐invasive blood pressure series, each consisting of three measurements from each monitor, were obtained from 20 neonates (birthweight 531–4660 g). Each measurement was compared with the invasive pressure. Results: Two factors appear to have a systematic effect on the difference between oscillometric and invasive pressure (the measurement deviance): the size of the infant, e.g., the arm circumference, and the monitor system. For small infants, the non‐invasively measured value tends to be too high. The deviance is partly reversed for larger infants (dependency on size significant for mean and diastolic pressure, p < 0.001). The difference between monitor systems is clearly significant (p < 0.001). Hewlett‐Packard™ gives the lowest pressure values for all pressures. Thus, Criticare™ and Dinamap™ tended to show values too high in the smallest infants, while Hewlett‐Packard™ tended to give values too low in larger infants. Birthweight, present weight and arm circumference affected measurement deviance approximately equally strongly, while factors such as the infant's sex, need of breath support and umbilical or radial arterial line were non‐significant. Conclusions: Blood pressure should preferably be measured invasively in severely ill neonates and preterm infants, being aware of pitfalls with measurements using different oscillometer monitors and the size/arm circumference of the infant.Keywords
This publication has 28 references indexed in Scilit:
- Non‐invasive oscillometric blood pressure measurement in very‐low‐birthweight infants: a comparison of two different monitor systemsActa Paediatrica, 1999
- Blood pressure ranges in premature infants. I. The first hours of lifeThe Journal of Pediatrics, 1994
- Randomized trial comparing dopamine and dobutamine in preterm infantsEuropean Journal of Pediatrics, 1993
- Non-invasive blood pressure monitoring in preterm infants receiving intensive careEuropean Journal of Pediatrics, 1992
- Dinamap fails to detect hypotension in very low birthweight infants.Archives of Disease in Childhood, 1986
- Is the blood pressure the same in both arms?Clinical Cardiology, 1985
- Blood pressure monitoring in neonates: Comparison of umbilical and peripheral artery catheter measurementsThe Journal of Pediatrics, 1984
- Noninvasive determination of systolic, diastolic and end-systolic blood pressure in neonates, infants and young children: Comparison with central aortic pressure measurementsThe American Journal of Cardiology, 1983
- Systolic pressure amplification in pedal arteries in childrenCritical Care Medicine, 1983
- EVALUATION OF AN AUTOMATIC OSCILLOMETRIC METHOD AND OF VARIOUS CUFFS FOR THE MEASUREMENT OF ARTERIAL PRESSURE IN THE NEONATEActa Paediatrica, 1982