RF‐induced heating for active implantable medical devices in dual parallel leads configurations at 1.5 T MRI
- 10 April 2023
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 90 (2) , 686-698
Abstract
Purpose: The Radiofrequency (RF)‐induced heating for an active implantable medical device (AIMD) with dual parallel leads is evaluated in this paper. The coupling effects between dual parallel leads are studied via simulations and experiments methods. The global transfer function technique is used to assess the RF‐induced heating for dual‐lead AIMDs inside four human body models.Methods: RF‐induced heating for spinal cord stimulator systems with 60 and 90 cm length leads are studied at three parallel dual‐lead configurations (closely spaced, 8 mm spaced, and 40 mm spaced) and a single‐lead configuration. The global transfer function method is used to develop the AIMD models of different configurations and is used for lead‐tip heating assessments inside human body models.Results: In simulation studies, the peak 1g specific absorption rate/temperatrue rises of dual parallel leads systems is lower than those from the single‐lead system. In experimental American Society for Testing and Materials phantom studies, the temperature rises for the single‐lead AIMD system can be 2.4 times higher than that from dual‐lead AIMD systems. For the spinal cord stimulator systems used in the study, the statistical analysis shows the RF‐induced heating of dual‐lead configurations are also lower than those from the single‐lead configuration inside all four human body models.Conclusion: For the AIMD system in this study, it shows that the coupling effects between the dual parallel leads of AIMD systems can reduce RF‐induced heating. The global transfer function for different spatial distance dual‐lead configurations can potentially provide a method for the RF‐induced heating evaluation for dual‐lead AIMD systems.Keywords
Funding Information
- Division of Industrial Innovation and Partnerships (1922389)
This publication has 28 references indexed in Scilit:
- Impact of capped and uncapped abandoned leads on the heating of an MR‐conditional pacemaker implantMagnetic Resonance in Medicine, 2014
- Historical and Present State of Neuromodulation in Chronic PainCurrent Pain and Headache Reports, 2013
- Computational and experimental studies of an orthopedic implant: MRI‐related heating at 1.5‐T/64‐MHz and 3‐T/128‐MHzJournal of Magnetic Resonance Imaging, 2012
- Role of the lead structure in MRI‐induced heating: In vitro measurements on 30 commercial pacemaker/defibrillator leadsMagnetic Resonance in Medicine, 2011
- A Generic Bioheat Transfer Thermal Model for a Perfused TissueJournal of Biomechanical Engineering, 2009
- Calculation of MRI‐induced heating of an implanted medical lead wire with an electric field transfer functionJournal of Magnetic Resonance Imaging, 2007
- Minimizing RF heating of conducting wires in MRIMagnetic Resonance in Medicine, 2007
- RF‐induced temperature elevation along metallic wires in clinical magnetic resonance imaging: Influence of diameter and lengthMagnetic Resonance in Medicine, 2004
- Dual-Chamber Pacing or Ventricular Backup Pacing in Patients With an Implantable DefibrillatorJAMA, 2002
- Shielded open-circuited sample holder for dielectric measurements of solids and liquidsIEEE Transactions on Instrumentation and Measurement, 1998