A semiautomated procedure for the determination of serum cholesterol using the abell‐kendall method

Authors:

Abstract
A semiautomated method for the measurement of serum total cholesterol, based on the Abell‐Kendall procedure, is described by which 100 to 140 samples may be analyzed in duplicate per day. The method was made possible by the development of a Liebermann color reagent which is stable for 8 hr. Mechanized pipetting units are used to measure all reagents and the 0.1 ml samples of serum. One analyst can analyze 60 samples in duplicate per day. Determinations are carried out on an assembly line basis so that 2 analysts can analyze more than twice as many samples as can one analyst. Representative data collected over a period of 2 years for the analysis of routine samples showed that the standard deviation for duplicates, randomized over 6‐day periods, ranged from 2.7 to 3.0 mg% after repeating the analysis of the 6 to 8% of samples whose duplicates differed by more than 9 mg%.