Anastomotic Troubleshooting
- 23 November 2021
- journal article
- review article
- Published by Georg Thieme Verlag KG in Clinics in Colon and Rectal Surgery
- Vol. 34 (06) , 385-390
Abstract
Anastomotic leak remains a critical and feared complication in colorectal surgery. The development of a leak can be catastrophic for a patient, resulting in overall increased morbidity and mortality. To help mitigate this risk, there are several ways to assess and potentially validate the integrity of a new anastomosis to give the patient the best chance of avoiding this postoperative complication. A majority of anastomoses will appear intact with no obvious sign of anastomotic dehiscence on gross examination. However, each anastomosis should be interrogated before the conclusion of an operation. The most common method to assess for an anastomotic leak is the air leak test (ALT). The ALT is a safe intraoperative method utilized to test the integrity of left-sided colon and rectal anastomoses and most importantly allows the ability to repair a failed test before concluding the operation. Additional troubleshooting is sometimes needed due to technical difficulties with the circular stapler. Problems, such as incomplete doughnuts and stapler misfiring, do occur and each surgeon should be prepared to address them. Publication Date: 23 November 2021 (online) © 2021. Thieme. All rights reserved. Thieme Medical Publishers, Inc. 333 Seventh Avenue, 18th Floor, New York, NY 10001, USAKeywords
This publication has 24 references indexed in Scilit:
- When staplers misfire: endoscopic rescue of low pelvic anastomosesTechniques in Coloproctology, 2010
- High Incidence of Technical Errors Involving the EEA Circular Stapler: A Single Institution ExperienceJournal of the American College of Surgeons, 2010
- Anastomotic leakage after colon cancer surgery: A predictor of significant morbidity and hospital mortality, and diminished tumour-free survivalEuropean Journal of Surgical Oncology, 2010
- Surgeons lack predictive accuracy for anastomotic leakage in gastrointestinal surgeryInternational Journal of Colorectal Disease, 2009
- Factors Associated With the Occurrence of Leaks in Stapled Rectal Anastomoses: A Review of 1,014 PatientsJournal of the American College of Surgeons, 1997
- Use of the circular stapler in 1000 consecutive colorectal anastomoses: Experience of one surgical teamSurgery, 1995
- Intraoperative air testing of colorectal anastomoses: A prospective, randomized trialBritish Journal of Surgery, 1990
- Stapled colorectal anastomoses: peroperative integrity of the anastomosis and risk of postoperative leakageInternational Journal of Colorectal Disease, 1986
- WHEN ARE EEA "STAPLER DOUGHNUTS" REALLY COMPLETE?The Lancet, 1980
- New Techniques of Gastrointestinal Anastomoses with the EEA StaplerAnnals of Surgery, 1979