AI-Assisted LVEF Assessment Using a Handheld Ultrasound Device: A Single-Center Comparative Study Against Cardiac Magnetic Resonance Imaging
Open Access
- 3 July 2025
- journal article
- research article
- Published by MDPI AG in Journal of Clinical Medicine
- Vol. 14 (13) , 4708
Abstract
Background/Objectives: Two-dimensional echocardiography (2D echo) is widely used for assessing left ventricular ejection fraction (LVEF). This single-center comparative study aims to evaluate the accuracy of LVEF measurements obtained using the AI-assisted handheld ultrasound device Kosmos against cardiac magnetic resonance (CMR), the current gold standard. Methods: A total of 49 adult patients undergoing clinically indicated CMR were prospectively enrolled. AI-based LVEF measurements were compared with CMR using the Wilcoxon signed-rank test, Pearson correlation, multivariable linear regression, and Bland–Altman analysis. All analyses were performed using STATA v18.0. Results: Median LVEF was 57% (CMR) vs. 55% (AI-Echo), with no significant difference (p = 0.51). Strong correlation (r = 0.99) and minimal bias (1.1%) were observed. Conclusions: The Kosmos AI-based autoEF algorithm demonstrated excellent agreement with CMR-derived LVEF values. Its speed and automation make it promising for bedside assessment in emergency departments, intensive care units, and outpatient clinics. This study aims to fill the gap in current clinical evidence by evaluating, for the first time, the agreement between LVEF measurements obtained via Kosmos’ AI-assisted autoEF and those from cardiac MRI (CMR), the gold standard for ventricular function assessment. This comparison is critical for validating the reliability of portable AI-driven echocardiographic tools in real-world clinical practice. However, these findings derive from a selected population at a single Italian center and should be validated in larger, diverse cohorts before assuming global generalizability.Keywords
This publication has 29 references indexed in Scilit:
- Point-of-care ultrasound to assess left ventricular ejection fraction in heart failure in unselected patients in primary care: a systematic reviewFamily Practice, 2024
- Multicenter validation study for automated left ventricular ejection fraction assessment using a handheld ultrasound with artificial intelligenceScientific Reports, 2024
- Diagnostic accuracy of point-of-care ultrasound with artificial intelligence-assisted assessment of left ventricular ejection fractionnpj Digital Medicine, 2023
- Improved assessment of left ventricular ejection fraction using artificial intelligence in echocardiography: A comparative analysis with cardiac magnetic resonance imagingInternational Journal of Cardiology, 2023
- Clinical validation of an artificial intelligence-assisted algorithm for automated quantification of left ventricular ejection fraction in real time by a novel handheld ultrasound deviceEuropean Heart Journal - Digital Health, 2022
- Handheld Point-of-Care Ultrasound Probes: The New Generation of POCUSJournal of Cardiothoracic and Vascular Anesthesia, 2020
- The use of handheld ultrasound devices: a position statement of the European Association of Cardiovascular Imaging (2018 update)European Heart Journal - Cardiovascular Imaging, 2018
- Machine learning for medical ultrasound: status, methods, and future opportunitiesAbdominal Radiology, 2018
- Recommendations for Cardiac Chamber Quantification by Echocardiography in Adults: An Update from the American Society of Echocardiography and the European Association of Cardiovascular ImagingJournal of the American Society of Echocardiography, 2015
- Recommendations for chamber quantification☆European Heart Journal - Cardiovascular Imaging, 2006