Abstract
A study was made with intra-operative flash--visual evoked potentials (VEP) monitored using a fibre-optic/contact lens photo stimulator in 57 patients undergoing intra-orbital surgical procedures with potential risk to the optic nerve. The VEPs recorded under enflurane and nitrous oxide anaesthesia did not differ significantly in latency or amplitude from the pre-operative recordings. Transient abolition of the VEP was seen under many circumstances and did not correlate with the outcome of surgery, but absence of a previously normal VEP for more than four minutes during surgical manipulation within the orbit did show a correlation with post operative impairment of vision. The technique provides early warning to the surgeon of threats to the integrity of the optic nerve.Keywords
This publication has 14 references indexed in Scilit:
- Factors that limit the use of flash visual evoked potentials for surgical monitoringElectroencephalography and Clinical Neurophysiology/Evoked Potentials Section, 1988
- Are Flash-evoked Visual Potentials Useful for Intraoperative Monitoring of Visual Pathway Function?Neurosurgery, 1987
- A contact lens photostimulator for surgical monitoringElectroencephalography and Clinical Neurophysiology, 1987
- Evoked Potential Monitoring in Neurosurgical Practice1986
- Cavitron ultrasonic surgical aspirator and visual evoked potential monitoring for chiasmal gliomas in childrenJournal of Neurosurgery, 1985
- The application of flash visual evoked potentials during operations on the anterior visual pathways*Neurological Research, 1985
- INTRAOPERATIVE MONITORING OF EVOKED POTENTIALSAnnals of the New York Academy of Sciences, 1982
- Monitoring optic nerve function during craniotomyJournal of Neurosurgery, 1976
- Neurophysiologic effects of general anesthetics. I. The electroencephalogram and sensory evoked responses in man.1973
- Continuous monitoring of the visually evoked response during intra-orbital surgery.1973