Ceramic–Titanium Alloy Artificial Hip Joint Wear Simulation and Experimental Study
Open Access
Abstract
The wear of artificial joints can lead to joint noise and tissue pathology within the human body, which is a primary factor affecting their service life. In response to the issue of wear in ceramic–titanium alloy artificial hip joints, this study employed hip joint wear simulations and experimental wear testing on hip joint specimens to investigate the impact of different contact surface parameters on the wear of ceramic–titanium alloy articulating surfaces. The objective was to provide guidance for joint surface treatment to minimize wear. The simulation results demonstrated that the contacting surfaces of the articulating components exhibited a crescent-shaped surface composition before and after wear. The initial variation in the surface friction coefficient had minimal influence on the wear rate after stabilization, whereas excessively high friction coefficients led to erratic changes in wear depth. Based on the simulation results, experimental research was conducted to compare the wear results of different surface roughness values ranging from 60 to 550 nm. The findings revealed that a surface roughness of 150 nm exhibited the least amount of wear and the best anti-wear performance. Furthermore, an exploration of the mechanism behind the influence of different surface friction coefficients on the wear of the articulating surfaces provided valuable insights for surface processing and wear analysis of artificial joints.Keywords
Funding Information
- National Natural Science Foundation of China (51575493, LGG19E050025)
- Natural Science Foundation of Zhejiang Province (51575493, LGG19E050025)
This publication has 36 references indexed in Scilit:
- Numerical investigation on particle impact erosion in ultrasonic-assisted abrasive slurry jet micro-machining of glassesPowder Technology, 2017
- Effect of roughness on the wear behavior of HVOF coatings dry sliding against a friction materialWear, 2016
- Numerical and experimental study on ultrasonic vibration-assisted micro-channelling of glasses using an abrasive slurry jetInternational Journal of Mechanical Sciences, 2016
- A review of squeaking in ceramic total hip prosthesesTribology International, 2016
- Comparison of the outcomes of three surgical treatments for end-stage temporomandibular joint diseaseInternational Journal of Oral and Maxillofacial Surgery, 2014
- Lubrication and wear modelling of artificial hip joints: A reviewTribology International, 2011
- Finite element analysis of the contact mechanics of ceramic-on-ceramic hip resurfacing prosthesesJournal of Bionic Engineering, 2010
- Development of computational wear simulation of metal-on-metal hip resurfacing replacementsJournal of Biomechanics, 2008
- Effect of swing phase load on metal-on-metal hip lubrication, friction and wearJournal of Biomechanics, 2005
- Student–Newman–Kuels controls the false discovery rateStatistics & Probability Letters, 2000