Effects of Sash Movement and Walk-bys on Aerodynamics and Contaminant Leakage of Laboratory Fume Cupboards
Open Access
- 1 January 2007
- journal article
- Published by National Institute of Industrial Health in Industrial Health
- Vol. 45 (2) , 199-208
Abstract
In order to speculate the physical mechanisms of contaminant leakage during sash movement and walk-bys through a laboratory fume cupboard, the complicated three-dimensional flow patterns and the real-time tracer gas leakage are studied via the laser-assisted flow visualization method and the standard gas sampling technique, respectively, over a transparent, full scale chemical fume cupboard. Through the flow visualization, the evolution of drastic changes of the flow pattern is demonstrated. The highly turbulent jet-like currents are induced by the unsteady flow motion near the cupboard face. Large-scale turbulent eddies accompanied with the jet-like currents obviously bring large amount of in-cupboard smoke out to the atmosphere. The turbulent mixing extends the size and the strength of the large-scale eddy circulations, and predominantly contributes to the mechanism that causes the severe spread of contaminant leakage in few seconds. The tracer gas tests that are conducted by employing pr-EN 14175:2003 method show consistent containment results with the flow visualization findings. The temporally evolving large-scale turbulent eddies induced by the sash movement and the walk-bys cause substantially high contaminant leakage to the environment and the breathing zone of the operator.Keywords
This publication has 10 references indexed in Scilit:
- Required response time for variable air volume fume hood controllersAnnals of Occupational Hygiene, 2000
- Containment testing of laboratory hoods in the as-used condition.Applied Occupational and Environmental Hygiene, 2000
- Fume Hood Performance: Using a Bypass in Variable Air Volume SystemsApplied Occupational and Environmental Hygiene, 1998
- Fume Hood Performance: Face Velocity Variability Inconsistent Air Volume SystemsApplied Occupational and Environmental Hygiene, 1998
- Application of a Tracer Gas Challenge with a Human Subject to Investigate Factors Affecting the Performance of Laboratory Fume HoodsAIHA Journal, 1998
- Reducing Employee Exposure Potential Using the ANSI/ASHRAE 110 Method of Testing Performance of Laboratory Fume Hoods as a Diagnostic ToolAIHA Journal, 1998
- A review of wake effects on worker exposureAnnals of Occupational Hygiene, 1995
- The Effect of Contaminant Source Momentum on a Workerʼs Breathing Zone Concentration in a Uniform FreestreamAIHA Journal, 1992
- Experimental and numerical investigations of the performance of fume cupboardsBuilding and Environment, 1991
- A New Method for Quantitative, In-Use Testing of Laboratory Fume HoodsAIHA Journal, 1989