Comparison of spin-echo echo-planar imaging magnetic resonance elastography with gradient-recalled echo magnetic resonance elastography and their correlation with transient elastography
Authors:
- Jin Woo YoonJYJin Woo Yoon
- Eun Sun LeeELEun Sun Lee
- Hyun Jeong ParkHPHyun Jeong Park
- Sung Bin ParkSung Bin Park
- Young Youn ChoYCYoung Youn Cho
- Stephan KannengiesserSKStephan Kannengiesser
- Erlangen Siemens Healthcare GmbHEGErlangen Siemens Healthcare GmbH
- Joonho HurJHJoonho Hur
- 1 July 2022
- journal article
- research article
- Published by Galenos Yayinevi in Diagnostic and Interventional Radiology
- Vol. 28 (4) , 294-300
Abstract
PURPOSE This study aimed to assess the agreement between liver stiffness (LS) values obtained by the gradient-recalled echo (GRE) magnetic resonance elastography (MRE) and spin-echo echo-planar imaging (SE-EPI) MRE with those of transient elastography ( TE), respectively. METHODS We retrospectively included 48 participants who underwent liver MRE with both GRE and SE-EPI sequences in the same session and also TE within 1 year. We obtained LS values for MRE by drawing free-hand region of interest, and TE was performed using a FibroScan device. We assessed the relationship between the mean LS values obtained by each MRE sequence and TE using the correlation coefficients and Bland-Altman plots, respectively. We also compared LS values and technical failure rates of measured values from MRE between SE-EPI and GRE sequences using the paired t-test and McNemar's test. The MRE failure was defined as the absence of pixel value with a confidence index above 95%. RESULTS The LS values from SE-EPI and GRE sequences strongly correlated with those from TE (GRE; r = 0.73, P <.001 vs. SE-EPI; r = 0.79, P <.001). In addition, the LS values from the 2 MRE sequences showed excellent relationship (intraclass correlation coefficient, 0.94 [0.89-0.97], P <.001). The LS values from SE-EPI and GRE MRE were not significantly different (4.14 kPa vs. 3.88 kPa, P =.19). Furthermore, the technical success rate of SE-EPI MRE was superior to that of GRE (100% vs. 83.8%, P =.031). CONCLUSION The measured LS values obtained using TE correlated strongly with those obtained using GRE and SE-EPI MRE techniques, even though SE-EPI-MRE resulted a higher technical success rate than GRE-MRE. Therefore, we believe that TE, GRE, and SE-EPI MR elastography techniques may complement each other according to the appropriate individual situation.Keywords
This publication has 38 references indexed in Scilit:
- Rapid acquisition of multifrequency, multislice and multidirectional MR elastography data with a fractionally encoded gradient echo sequenceNMR in Biomedicine, 2013
- Magnetic resonance elastography of liver: Technique, analysis, and clinical applicationsJournal of Magnetic Resonance Imaging, 2013
- Interobserver concordance in the assessment of liver fibrosis in HIV/HCV-coinfected patients using transient elastometryEuropean Journal of Gastroenterology & Hepatology, 2010
- Transient elastography in chronic hepatitis B: An Asian perspectiveWorld Journal of Gastroenterology, 2010
- Cross‐validation of magnetic resonance elastography and ultrasound‐based transient elastography: A preliminary phantom studyJournal of Magnetic Resonance Imaging, 2009
- Magnetic Resonance Elastography for the Noninvasive Staging of Liver FibrosisGastroenterology, 2008
- Models of liver fibrosis: exploring the dynamic nature of inflammation and repair in a solid organJournal of Clinical Investigation, 2007
- Reproducibility of transient elastography in the evaluation of liver fibrosis in patients with chronic liver diseaseGut, 2007
- Sources of variability in histological scoring of chronic viral hepatitis†Hepatology, 2005
- Sampling Variability of Liver Fibrosis in Chronic Hepatitis CHepatology, 2003