Abstract
A contactless method of probing and analyzing multifunctional oxide interfaces using an electron energy analyzer inside a scanning electron microscope is presented. High contrast experimental secondary electron analyzer signals are used to detect changes in the interface conductivity of a LaAlO3/SrTiO3 sample. Monte Carlo simulations of the primary beam/specimen interaction are carried out and correlated with the experimental results in order to help understand the role of the primary beam energy and adjust it to enhance contrast.

This publication has 12 references indexed in Scilit: