Abstract
Objectives: To determine whether X‐ray fused with MRI (XFM) is beneficial for select transcatheter congenital heart disease interventions.Background: Complex transcatheter interventions often require three‐dimensional (3D) soft tissue imaging guidance. Fusion imaging with live X‐ray fluoroscopy can potentially improve and simplify procedures.Methods: Patients referred for select congenital heart disease interventions were prospectively enrolled. Cardiac MRI data was overlaid on live fluoroscopy for procedural guidance. Likert scale operator assessments of value were recorded. Fluoroscopy time, radiation exposure, contrast dose, and procedure time were compared to matched cases from our institutional experience.Results: Forty‐six patients were enrolled. Pre‐catheterization, same day cardiac MRI findings indicated intervention should be deferred in nine patients. XFM‐guided cardiac catheterization was performed in 37 (median age 8.7 years [0.5–63 years]; median weight 28 kg [5.6–110 kg]) with the following prespecified indications: pulmonary artery (PA) stenosis (n = 13), aortic coarctation (n = 12), conduit stenosis/insufficiency (n = 9), and ventricular septal defect (n = 3). Diagnostic catheterization showed intervention was not indicated in 12 additional cases. XFM‐guided intervention was performed in the remaining 25. Fluoroscopy time was shorter for XFM‐guided intervention cases compared to matched controls. There was no significant difference in radiation dose area product, contrast volume, or procedure time. Operator Likert scores indicated XFM provided useful soft tissue guidance in all cases and was never misleading.Conclusions: XFM provides operators with meaningful three‐dimensional soft tissue data and reduces fluoroscopy time in select congenital heart disease interventions.Keywords
Funding Information
- National Heart Lung and Blood Institute (Z01-HL006039, HHSN268201500001C)
- Siemens Medical (CNMC-2013-AX-01-02-C213599)
This publication has 23 references indexed in Scilit:
- Usefulness of echocardiographic-fluoroscopic fusion imaging in children with congenital heart diseaseArchives of Cardiovascular Diseases, 2018
- Novel Three-Dimensional Image Fusion Software to Facilitate Guidance of Complex Cardiac CatheterizationPediatric Cardiology, 2017
- Real‐time MRI guidance of cardiac interventionsJournal of Magnetic Resonance Imaging, 2017
- Feasibility and Safety of Using a Fused Echocardiography/Fluoroscopy Imaging System in Patients with Congenital Heart DiseaseJournal of the American Society of Echocardiography, 2016
- Optimized protocols for cardiac magnetic resonance imaging in patients with thoracic metallic implantsPediatric Radiology, 2015
- X‐ray magnetic resonance fusion modality may reduce radiation exposure and contrast dose in diagnostic cardiac catheterization of congenital heart diseaseCatheterization and Cardiovascular Interventions, 2014
- Multimodality 3D‐roadmap for cardiovascular interventions in congenital heart disease—A single‐center, retrospective analysis of 78 casesCatheterization and Cardiovascular Interventions, 2012
- The use of three‐dimensional rotational angiography to assess the pulmonary circulation following cavo‐pulmonary connection in patients with single ventricleCatheterization and Cardiovascular Interventions, 2012
- Technology preview: X‐ray fused with magnetic resonance during invasive cardiovascular proceduresCatheterization and Cardiovascular Interventions, 2007
- Cancer risks following diagnostic and therapeutic radiation exposure in childrenPediatric Radiology, 2006