Active cancellation of stray magnetic fields in a Bose-Einstein condensation experiment
Authors:
- 1 February 2007
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 78 (2) , 024703
Abstract
A method of active field cancellation is described, which greatly reduces the stray magnetic field within the trap region of a Bose-Einstein condensation experiment. An array of six single-axis magnetic sensors is used to interpolate the field at the trap center, thus avoiding the impractical requirement of placing the sensor within the trap. The system actively suppresses all frequencies from dc to approximately 3000 Hz, and the performance is superior to conventional active Helmholtz cancellation systems. A method of reducing the field gradient, by driving the six Helmholtz coils independently, is also investigated.Keywords
This publication has 11 references indexed in Scilit:
- Bose–Einstein condensation of metastable helium in a bi-planar quadrupole Ioffe configuration trapOptics Communications, 2007
- All-Optical Formation of an Atomic Bose-Einstein CondensatePhysical Review Letters, 2001
- Atom Laser with a cw Output CouplerPhysical Review Letters, 1999
- Bose-Einstein Condensation in a Gas of Sodium AtomsPhysical Review Letters, 1995
- Evidence of Bose-Einstein Condensation in an Atomic Gas with Attractive InteractionsPhysical Review Letters, 1995
- Observation of Bose-Einstein Condensation in a Dilute Atomic VaporScience, 1995
- Conventional magnetic shieldingJournal of Physics D: Applied Physics, 1987
- SQUID instruments and shielding for low-level magnetic measurementsJournal of Applied Physics, 1977
- Magnetic shieldsIEEE Transactions on Magnetics, 1970
- Use of an Operational Amplifier with Helmholtz Coils for Reducing ac Induced Magnetic FieldsReview of Scientific Instruments, 1961