Leadless pacing: Going for the jugular

Authors:

Abstract
Background Leadless pacing is generally performed from a femoral approach. However, the femoral route is not always available. Until now, data regarding implantation using a jugular approach other than a single case report is lacking. Methods Case records of all patients who underwent internal jugular venous (IJV) leadless pacemaker implantation (Micra, Medtronic) at our center were analysed retrospectively. Results 19 patients underwent IJV leadless pacemaker implantation, 9 female, mean age 77.5 ±9.6 yrs; permanent AF in all patients with normal left ventricular ejection fraction. Implant indication was AV conduction disturbance in 10, pre AV node ablation in 7 and replacement of a conventional VVI pacemaker in 2 (infection in 1 and lead malfunction in the other). The device was positioned at the superior septum in 7 patients, apicoseptal in 7 patients and midseptal in 5 patients. In 12 patients, a sufficient device position was obtained at the first attempt, in 3 at the 2nd, in 1 at the 3rd, in 1 at the 4th and in 2 at 6th attempt. The mean pacing threshold was 0.56± 0.39V at 0.24 ms pulse width, sensed amplitude was 9.1±3.2 mV , mean fluoroscopy duration 3.1±1.6 min. There were no vascular or other complications. At follow‐up electrical parameters remained stable in 18 of 19 patients. Conclusion Although experience is minimal, we suggest that the IJV approach is safe and may be considered in patients where the femoral approach is contraindicated. This article is protected by copyright. All rights reserved