Agreement between two photoplethysmography-based wearable devices for monitoring heart rate during different physical activity situations: a new analysis methodology
Authors:
Open Access
- 14 September 2022
- journal article
- research article
- Published by Springer Nature in Scientific Reports
- Vol. 12 (1) , 1-11
Abstract
Wearables are being increasingly used to monitor heart rate (HR). However, their usefulness for analyzing continuous HR in research or at clinical level is questionable. The aim of this study is to analyze the level of agreement between different wearables in the measurement of HR based on photoplethysmography, according to different body positions and physical activity levels, and compared to a gold-standard ECG. The proposed method measures agreement among several time scales since different wearables obtain HR at different sampling rates. Eighteen university students (10 men, 8 women; 22 ± 2.45 years old) participated in a laboratory study. Participants simultaneously wore an Apple Watch and a Polar Vantage watch. ECG was measured using a BIOPAC system. HR was recorded continuously and simultaneously by the three devices, for consecutive 5-min periods in 4 different situations: lying supine, sitting, standing and walking at 4 km/h on a treadmill. HR estimations were obtained with the maximum precision offered by the software of each device and compared by averaging in several time scales, since the wearables obtained HR at different sampling rates, although results are more detailed for 5 s and 30 s epochs. Bland–Altman (B-A) plots show that there is no noticeable difference between data from the ECG and any of the smartwatches while participants were lying down. In this position, the bias is low when averaging in both 5 s and 30 s. Differently, B-A plots show that there are differences when the situation involves some level of physical activity, especially for shorter epochs. That is, the discrepancy between devices and the ECG was greater when walking on the treadmill and during short time scales. The device showing the biggest discrepancy was the Polar Watch, and the one with the best results was the Apple Watch. We conclude that photoplethysmography-based wearable devices are suitable for monitoring HR averages at regular intervals, especially at rest, but their feasibility is debatable for a continuous analysis of HR for research or clinical purposes, especially when involving some level of physical activity. An important contribution of this work is a new methodology to synchronize and measure the agreement against a gold standard of two or more devices measuring HR at different and not necessarily even paces.Keywords
Funding Information
- Ministerio de Ciencia e Innovación (PID2019-107473RB-C21, PID2019-107473RB-C22, PID2019-107473RB-C21, PID2019-107473RB-C22, PID2019-107473RB-C22, PID2019-107473RB-C21)
This publication has 31 references indexed in Scilit:
- Measures of sleep and cardiac functioning during sleep using a multi-sensory commercially-available wristband in adolescentsPhysiology & Behavior, 2016
- A methodology to quantify the differences between alternative methods of heart rate variability measurementPhysiological Measurement, 2015
- Validation of photoplethysmography as a method to detect heart rate during rest and exerciseJournal of Medical Engineering & Technology, 2015
- Validity of the Smarthealth Watch to Measure Heart Rate During Rest and ExerciseMeasurement in Physical Education and Exercise Science, 2011
- Comparison of Omega Wave System and Polar S810i to Detect R-R Intervals at RestInternational Journal of Sports Medicine, 2010
- Comparison of Polar 810 s and an Ambulatory ECG System for RR Interval Measurement During Progressive ExerciseInternational Journal of Sports Medicine, 2004
- Validation of a cardiac monitor for measuring heart rate variability in adult female pigs: accuracy, artefacts and editingPhysiology & Behavior, 2004
- Scale Invariance in the Nonstationarity of Human Heart RatePhysical Review Letters, 2001
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986