Effect of standard cuff on blood pressure readings in patients with obese arms. How frequent are arms of a ‘large circumference’?
Authors:
- Salvador Fonseca-ReyesSFSalvador Fonseca-Reyes
- Javier García de Alba-GarcíaJAJavier García de Alba-García
- José Z. Parra-CarrilloJPJosé Z. Parra-Carrillo
- José Antonio Paczka-ZapataJPJosé Antonio Paczka-Zapata
- 1 June 2003
- journal article
- clinical trial
- Published by Wolters Kluwer Health in Blood Pressure Monitoring
- Vol. 8 (3) , 101-106
Abstract
To measure the effect on blood pressure readings when a standard cuff is used on patients with arms of a large circumference, and to determine the frequency of arms of a large circumference. Blood pressures were taken in 120 subjects with an arm circumference greater than 33 cm. Also, the arm circumference was determined in 244 patients from a family health unit, and in 216 patients from a hypertension clinic. A mercury sphygmomanometer and two different cuff sizes were used in a random sequence; therefore, 60 patients' blood pressure were first measured with a large cuff, followed by a standard cuff; the opposite sequence was then applied for another 60 patients. With the obtained values and using a regression analysis, the difference in blood pressure overestimation was calculated. Arm circumference measurement percentages were used to determine the frequency of arms of a large circumference. Both systolic and diastolic blood pressures were significantly greater when the standard cuff was used. For every 5 cm increase in arm circumference, starting at 35 cm, a 2-5 mmHg increase in systolic blood pressure, and a 1-3 mmHg increase in diastolic blood pressure was observed. The prevalence of arms with a large circumference in the family medicine unit and hypertension clinic was 42% and 41.8%, respectively. There is an overestimation of blood pressure when a standard cuff is used in obese subjects. The high prevalence of these individuals in our environment, both in the hypertensive and normotensive population, makes it necessary to have on hand different sizes of cuffs for taking blood pressure in order to avoid incorrect decisions.Keywords
This publication has 12 references indexed in Scilit:
- Blood pressure measurementBMJ, 2001
- Prevalence of blood pressure cuff sizes in a referral practice of 430 consecutive adult hypertensivesBlood Pressure Monitoring, 2001
- The Spread of the Obesity Epidemic in the United States, 1991-1998JAMA, 1999
- Knowledge of blood pressure measurement among a teaching hospital staff in a developing nationJournal of Human Hypertension, 1997
- Evaluation of the Technique Used by Health-Care Workers for Taking Blood PressureHypertension, 1995
- Human blood pressure determination by sphygmomanometry.Circulation, 1993
- The Effect of Cuff Size on Blood Pressure Measurement in AdultsEpidemiology, 1991
- Ratio of cuff width/arm circumference as a determinant of arterial blood pressure measurements in adultsJournal of Internal Medicine, 1990
- Miscuffing: inappropriate blood pressure cuff application.Circulation, 1983
- ERROR IN BLOOD-PRESSURE MEASUREMENT DUE TO INCORRECT CUFF SIZE IN OBESE PATIENTSThe Lancet, 1982