Transcutaneous transmission of photobiomodulation light to the spinal canal of dog as measured from cadaver dogs using a multi-channel intra-spinal probe
- 16 March 2019
- journal article
- research article
- Published by Springer Nature in Lasers in Medical Science
- Vol. 34 (8) , 1645-1654
Abstract
The target level photobiomodulation (PBM) irradiances along the thoracic to lumbar segment of the interior spinal canal in six cadaver dogs resulting from surface illumination at 980 nm were measured. Following a lateral hemi-laminectomy, a flexible probe fabricated on a plastic tubular substrate of 6.325 mm diameter incorporating nine miniature photodetectors was embedded in the thoracic to lumbar segment of the spinal canal. Intra-spinal irradiances at the nine photodetector sites, spanning an approximate 8 cm length caudal to T13, were measured for various applied powers of continuous wave (CW) surface illumination at 980 nm with a maximal power of 10 W corresponding to a surface irradiance of 3.14 W/cm2. The surface illumination conditions differed in skin transmission when the probe was off-contact with tissue and probe-skin contact when the skin was in place. For each condition of surface illumination, the beam was directed to respectively T13 (surface site 1), a spinal column site 4 cm caudal to T13 (surface site 5), and a spinal column site 8 cm caudal to T13 (surface site 9). Off-contact surface irradiation of 3.14 W/cm2 at surface sites 1, 5, and 9 transmitted respectively 234.0 ± 120.7 μW/cm2, 230.7 ± 178.3 μW/cm2, and 130.2 ± 169.6 μW/cm2 to the spinal canal without the skin, and respectively 35.7 ± 33.2 μW/cm2, 50.9 ± 75.3 μW/cm2, and 15.7 ± 16.3 μW/cm2 with the skin. Transmission with skin was as low as 12% of the transmission without the skin. On-contact surface irradiation of 3.14 W/cm2 at surface sites 1, 5, and 9 transmitted respectively 44.6 ± 43.1 μW/cm2, 85.4 ± 139.1 μW/cm2, and 22.0 ± 23.6 μW/cm2 to the spinal canal. On-contact application increased transmission by a maximum of 67% comparing to off-contact application. The information gathered highlights the need to clinically consider the impact of skin transmission and on-contact application technique when attempting to treat spinal cord disease with PBM.Keywords
Funding Information
- LiteCure LLC (Research Grant)
This publication has 29 references indexed in Scilit:
- Transcranial near-infrared photobiomodulation attenuates memory impairment and hippocampal oxidative stress in sleep-deprived miceBrain Research, 2018
- Transcranial low-level laser therapy improves brain mitochondrial function and cognitive impairment in D-galactose–induced aging miceNeurobiology of Aging, 2017
- Use of low level laser therapy to control neuropathic pain: A systematic reviewJournal of Photochemistry and Photobiology B: Biology, 2016
- Assessment of feasibility and efficacy of Class IV laser therapy for postoperative pain relief in off-pump coronary artery bypass surgery patients: A pilot studyAnnals of Cardiac Anaesthesia, 2015
- Restoring function after spinal cord injury: towards clinical translation of experimental strategiesThe Lancet Neurology, 2014
- Low-level laser therapy for spinal cord injury in rats: effects of polarizationJournal of Biomedical Optics, 2013
- Avaliação do papel de opioides endógenos na analgesia do laser de baixa potência, 820 nm, em joelho de ratos WistarRevista Dor, 2012
- Influência da naloxona e metisergida sobre o efeito analgésico do laser em baixa intensidade em modelo experimental de dorRevista Brasileira de Anestesiologia, 2010
- 810 nm Wavelength light: An effective therapy for transected or contused rat spinal cordLasers in Surgery and Medicine, 2009
- Light promotes regeneration and functional recovery and alters the immune response after spinal cord injuryLasers in Surgery and Medicine, 2005