Performance Monitoring of Mycobacterium tuberculosis Dried Culture Spots for Use with the GeneXpert System within a National Program in South Africa
Authors:
Open Access
- 1 December 2013
- journal article
- research article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 51 (12) , 4018-4021
Abstract
The use of dried culture spots (DCSs) has been reported in the verification of GeneXpert instruments as being “fit for purpose” for the South African National implementation program. We investigated and compared the performance of the DCSs for verification across different bulk batches, testing the settings and cadre of staff, and the Xpert MTB/RIF assay version. Four bulk batches (V005 to V008) were used to prepare (i) 619 DCS panels for laboratory testing on G3 or G4 cartridges by a technologist, (ii) 13 DCS panels (batch V005) used for clinic verification on G3 cartridges by a nurse or lay counselor, and (iii) 20 DCS panels (batch V005) used for the verification of 10 GeneXpert 16 module instruments in mobile vehicles on the G3 cartridge performed by a scientist. The stabilities of the DCSs over 6 months at 4°C, room temperature, and 37°C were investigated. The mean cycle threshold ( C T ) and standard deviation (SD) for probe A were calculated. The proportions of variability in the C T values across bulk batches, assay versions, and settings and cadre of staff were determined using regression analysis. Overall, the DCSs demonstrated SDs of 3.3 ( n = 660) for the G3 cartridges and 3.8 ( n = 1,888) for the G4 cartridges, with an overall error rate of 1.5% and false rifampin resistance rate of 0.1%. The proportions of variability ( R 2 ) in the C T values explained by batch were 14%, by setting and cadre of staff, 5.6%, and by assay version, 4.2%. The most stable temperature in a period of up to 6 months was 37°C (SD, 2.7). The DCS is a robust product suitable for storage, transport, and use at room temperature for the verification of the GeneXpert instrument, and the testing can be performed by non-laboratory-trained personnel in nonlaboratory settings.Keywords
This publication has 5 references indexed in Scilit:
- Scaling up Xpert MTB/RIF technology: the costs of laboratory‐ vs. clinic‐based roll‐out in South AfricaTropical Medicine & International Health, 2012
- Dried Culture Spots for Xpert MTB/RIF External Quality Assessment: Results of a Phase 1 Pilot Study in South AfricaJournal of Clinical Microbiology, 2011
- Rapid Molecular Detection of Tuberculosis and Rifampin ResistanceNew England Journal of Medicine, 2010
- Evaluation of the Analytical Performance of the Xpert MTB/RIF AssayJournal of Clinical Microbiology, 2010
- The resuscitation‐promoting factors of Mycobacterium tuberculosis are required for virulence and resuscitation from dormancy but are collectively dispensable for growth in vitroMolecular Microbiology, 2007