Intraocular lens power formula based on vergence calculation and lens design
Author:
- 1 October 1997
- journal article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 23 (8) , 1200-1207
Abstract
To present a new method of intraocular lens (IOL) power calculation that takes into account the power-dependent variation in lens design. Department of Ophthalmology, Aalborg, Denmark. Information on exact IOL design is derived from the manufacturers' cutting cards. These data were built into an IOL calculation formula based on exact ray tracing and vergence calculation. All algorithms are demonstrated. The method is optically correct as all refractive surfaces are characterized with respect to both position and refractive power. In simulation models, the Naeser and the Holladay formulas performed similarly, while the SRK/T formula predicted higher postoperative refractions for low-power IOLs. It is possible to incorporate the exact IOL design into an IOL power calculation formula. Theoretically, the Naeser formula should increase the accuracy of IOL power calculation; however, this has yet to be proved from empirical data. The formula provides an advantage in analysis of postoperative pseudophakia for experimental/scientific purposes because all intraocular distances and powers may be measured or calculated.Keywords
This publication has 12 references indexed in Scilit:
- An improved universal theoretical formula for intraocular lens power predictionJournal of Cataract & Refractive Surgery, 1993
- Calculation of the thickness of an intraocular lensJournal of Cataract & Refractive Surgery, 1993
- The Vergence-based, Empirically Modified Intraocular Lens EquationEuropean journal of Implant and Refractive Surgery, 1991
- Vaulted posterior chamber lenses and the posterior capsuleJournal of Cataract & Refractive Surgery, 1990
- Comparison of the SRK/T formula and other theoretical and regression formulasJournal of Cataract & Refractive Surgery, 1990
- Development of the SRK/T intraocular lens implant power calculation formulaJournal of Cataract & Refractive Surgery, 1990
- Biometry of the posterior lens capsule: A new method to predict pseudophakic anterior chamber depthJournal of Cataract & Refractive Surgery, 1990
- A three-part system for refining intraocular lens power calculationsJournal of Cataract & Refractive Surgery, 1988
- Theoretical approach to intraocular lens calculation using Gaussian opticsJournal of Cataract & Refractive Surgery, 1987
- Posterior chamber implant power calculation: regression formulasAmerican Intra-Ocular Implant Society Journal, 1980