Changes in Pediatric-Sized Endotracheal Tube Cuff Pressure With Elevation Gain
Authors:
- 1 August 2018
- journal article
- research article
- Published by Wolters Kluwer Health in Pediatric Emergency Care
- Vol. 34 (8) , 570-573
Abstract
Objectives: Prolonged excessive endotracheal cuff pressure greater than 30 cmH2O is thought to cause ischemic airway injury. Excessive cuff pressure with altitude gain during air medical transport has been previously described in adult patients. It is poorly understood how pediatric-sized endotracheal tube (ETT) cuffs behave with atmospheric pressure change during flight. Methods: In ex vivo models 4.0, 6.0, and 8.0, ETTs restricted within scaled syringe tubing were inflated to 20 cmH2O. Pressure was measured against 1500 ft elevation gain in ground and flight models. In an in vivo observation of pediatric patient transport, change in cuff pressure was measured between takeoff and helicopter peak flight altitudes. Results: In the ex vivo ground model, endotracheal cuff pressure increased linearly with altitude and exceeded 40 cmH2O in all tube sizes. Comparable pressure change was demonstrated in the flight model. No difference was demonstrated in the degree of pressure change between ETT sizes. In the in vivo observations during patient transport, pressure increase was consistent with that seen in the ex vivo models. Conclusions: Children who are intubated with cuffed ETTs for air medical transport are subject to excessive endotracheal cuff pressure at even low flight altitudes. Endotracheal tube size did not affect the degree of cuff pressure change, contrary to previous study. These findings need to be validated and correlated to patient clinical outcomes. The implications of these data need to be considered clinically particularly for prolonged transport of intubated pediatric patients at elevation.Keywords
This publication has 11 references indexed in Scilit:
- Managing endotracheal tube cuff pressure at altitudeJournal of Trauma and Acute Care Surgery, 2014
- Continuous real time endotracheal tube cuff pressure waveformJournal of Clinical Monitoring and Computing, 2014
- Endotracheal Tube Cuff Pressure Monitoring: A Review of the EvidenceJournal of Perioperative Practice, 2011
- Endotracheal Tube and Laryngeal Mask Airway Cuff Pressures Can Exceed Critical Values During Ascent to Higher AltitudePediatric Emergency Care, 2011
- Endotracheal Tube Intracuff Pressure During Helicopter TransportAnnals of Emergency Medicine, 2010
- Endotracheal tube and laryngeal mask airway cuff volume changes with altitude: a rule of thumb for aeromedical transportEmergency Medicine Journal, 2007
- Pressures within air‐filled tracheal cuffs at altitude – an in vivo studyAnaesthesia, 2004
- Pressure in the cuffs of tracheal tubes at altitudeAnaesthesia, 2002
- Dimensions of the growing trachea related to age and genderAmerican Journal of Roentgenology, 1986
- Endotracheal cuff pressure and tracheal mucosal blood flow: endoscopic study of effects of four large volume cuffs.BMJ, 1984