Correction Approach for Delta Function Convolution Model Fitting of Fluorescence Decay Data in the Case of a Monoexponential Reference Fluorophore
Open Access
- 11 June 2015
- journal article
- research article
- Published by Springer Nature in Journal of Fluorescence
- Vol. 25 (5) , 1169-1182
Abstract
A correction is proposed to the Delta function convolution method (DFCM) for fitting a multiexponential decay model to time-resolved fluorescence decay data using a monoexponential reference fluorophore. A theoretical analysis of the discretised DFCM multiexponential decay function shows the presence an extra exponential decay term with the same lifetime as the reference fluorophore that we denote as the residual reference component. This extra decay component arises as a result of the discretised convolution of one of the two terms in the modified model function required by the DFCM. The effect of the residual reference component becomes more pronounced when the fluorescence lifetime of the reference is longer than all of the individual components of the specimen under inspection and when the temporal sampling interval is not negligible compared to the quantity (τR −1 – τ−1)−1, where τR and τ are the fluorescence lifetimes of the reference and the specimen respectively. It is shown that the unwanted residual reference component results in systematic errors when fitting simulated data and that these errors are not present when the proposed correction is applied. The correction is also verified using real data obtained from experiment.Keywords
This publication has 33 references indexed in Scilit:
- Multiphoton Multispectral Fluorescence Lifetime Tomography for the Evaluation of Basal Cell CarcinomasPLOS ONE, 2012
- Fluorescence Instrument Response Standards in Two-Photon Time-Resolved SpectroscopyApplied Spectroscopy, 2010
- Fluorescence Lifetime Measurements and Biological ImagingChemical Reviews, 2010
- Förster resonance energy transfer (FRET)-based picosecond lifetime reference for instrument response evaluationMeasurement Science and Technology, 2009
- Spectral fluorescence lifetime detection and selective melanin imaging by multiphoton laser tomography for melanoma diagnosisExperimental Dermatology, 2009
- Combined non‐linear laser imaging (two‐photon excitation fluorescence microscopy, fluorescence lifetime imaging microscopy, multispectral multiphoton microscopy) in cutaneous tumours: first experiencesJournal of the European Academy of Dermatology and Venereology, 2009
- A compact, multidimensional spectrofluorometer exploiting supercontinuum generationJournal of Biophotonics, 2008
- In vivo multiphoton microscopy of NADH and FAD redox states, fluorescence lifetimes, and cellular morphology in precancerous epitheliaProceedings of the National Academy of Sciences of the United States of America, 2007
- Luminescence lifetime measurements. Elimination of phototube time shifts with the phase plane methodAnalytical Chemistry, 1985
- Evaluation of nanosecond pulse–fluorometry measurements—no need for the excitation functionReview of Scientific Instruments, 1981