Validation of Attitude and Heading Reference System and Microsoft Kinect for Continuous Measurement of Cervical Range of Motion Compared to the Optical Motion Capture System
Open Access
- 1 January 2016
- journal article
- Published by Korean Academy of Rehabilitation Medicine in Annals of Rehabilitation Medicine
- Vol. 40 (4) , 568-574
Abstract
To compare optical motion capture system (MoCap), attitude and heading reference system (AHRS) sensor, and Microsoft Kinect for the continuous measurement of cervical range of motion (ROM). Fifteen healthy adult subjects were asked to sit in front of the Kinect camera with optical markers and AHRS sensors attached to the body in a room equipped with optical motion capture camera. Subjects were instructed to independently perform axial rotation followed by flexion/extension and lateral bending. Each movement was repeated 5 times while being measured simultaneously with 3 devices. Using the MoCap system as the gold standard, the validity of AHRS and Kinect for measurement of cervical ROM was assessed by calculating correlation coefficient and Bland-Altman plot with 95% limits of agreement (LoA). MoCap and ARHS showed fair agreement (95% LoA10°) for measuring ROM in all directions. Intraclass correlation coefficient (ICC) values between MoCap and AHRS in -40° to 40° range were excellent for flexion/extension and lateral bending (ICC>0.9). ICC values were also fair for axial rotation (ICC>0.8). ICC values between MoCap and Kinect system in -40° to 40° range were fair for all motions. Our study showed feasibility of using AHRS to measure cervical ROM during continuous motion with an acceptable range of error. AHRS and Kinect system can also be used for continuous monitoring of flexion/extension and lateral bending in ordinary range.Keywords
This publication has 27 references indexed in Scilit:
- Inertial Measures of Motion for Clinical Biomechanics: Comparative Assessment of Accuracy under Controlled Conditions - Effect of VelocityPLOS ONE, 2013
- Measurement of Intervertebral Cervical Motion by Means of Dynamic X-Ray Image Processing and Data InterpolationInternational Journal of Biomedical Imaging, 2013
- Cervical Coupling Motion Characteristics in Healthy People Using a Wireless Inertial Measurement UnitEvidence-Based Complementary and Alternative Medicine, 2013
- Three-dimensional analysis of cervical spine segmental motion in rotationArchives of Medical Science, 2013
- Rating scales for cervical dystonia: a critical evaluation of tools for outcome assessment of botulinum toxin therapyJournal of Neural Transmission, 2012
- Accuracy and Resolution of Kinect Depth Data for Indoor Mapping ApplicationsSensors, 2012
- Validation of a Noninvasive Technique to Precisely Measure In Vivo Three-Dimensional Cervical Spine MovementSpine, 2011
- An integrated GPS/MEMS-IMU navigation system for an autonomous helicopterAerospace Science and Technology, 2006
- Real-time tracking of vertebral body movement with implantable reference microsensorsComputer Aided Surgery, 2006
- Inclination measurement of human movement using a 3-D accelerometer with autocalibrationIEEE Transactions on Neural Systems and Rehabilitation Engineering, 2004