Exhausted implanted pulse generator in sacral nerve stimulation for faecal incontinence: What next in daily practice for patients?
- 10 November 2015
- journal article
- research article
- Published by Springer Nature in International Journal of Colorectal Disease
- Vol. 31 (2) , 439-444
Abstract
Purpose The efficacy of sacral nerve stimulation in faecal incontinence relies on an implanted pulse generator known to have a limited lifespan. The long-term use of sacral nerve stimulation raises concerns about the true lifespan of generators. The aim of the study was to assess the lifespan of sacral nerve stimulation implanted pulse generators in daily practice, and the outcome of exhausted generator replacement, in faecal incontinent patients. Methods Faecal incontinent patients with pulse generators (Medtronic Interstim™ or InterstimII™) implanted in a single centre from 2001 to 2014 were prospectively followed up. Generator lifespan was measured according to the Kaplan-Meier method. Patients with a generator explanted/turned off before exhaustion were excluded. Morbidity of exhausted generator replacement and the outcome (Cleveland Clinic Florida Faecal Incontinence (CCF-FI) and Faecal Incontinence Quality of Life (FIQL) scores) were recorded. Results Of 135 patients with an implanted pulse generator, 112 (InterstimII™ 66) were included. Mean follow-up was 4.9 ± 2.8 years. The generator reached exhaustion in 29 (26 %) cases. Overall median lifespan of an implanted pulse generator was approximately 9 years (95 % CI 8–9.2). Interstim™ and InterstimII™ 25th percentile lifespan was 7.2 (CI 6.4–8.3) and 5 (CI 4–not reached) years, respectively. After exhaustion, generators were replaced, left in place or explanted in 23, 2 and 4 patients, respectively. Generator replacement was virtually uneventful. CCF-FI/FIQL scores remained unchanged after generator replacement (CCF-FI 8 ± 2 vs 7 ± 3; FIQL 3 ± 0.6 vs 3 ± 0.5; p = ns). Conclusion In this study, the implanted pulse generator observed median lifespan was 9 years. After exhaustion, generators were safely and efficiently replaced. The study also gives insight into long-term needs and costs of sacral nerve stimulation (SNS) therapy.Keywords
Funding Information
- Centre d’études et de recherches en chirurgie” (CEREC) of the University Hospital of Nantes
This publication has 21 references indexed in Scilit:
- Long-term outcomes of sacral nerve stimulation for faecal incontinenceBritish Journal of Surgery, 2015
- Resource implications of running a sacral neuromodulation service: a 10‐year experienceColorectal Disease, 2014
- Sacral neuromodulation: long-term outcome and quality of life in patients with faecal incontinenceColorectal Disease, 2010
- Sacral neuromodulation with implanted devices for urinary storage and voiding dysfunction in adultsCochrane Database of Systematic Reviews, 2009
- Sacral Nerve Stimulation is more Effective than Optimal Medical Therapy for Severe Fecal Incontinence: A Randomized, Controlled StudyDiseases of the Colon & Rectum, 2008
- Outcome and cost analysis of sacral nerve stimulation for faecal incontinenceBritish Journal of Surgery, 2006
- Sacral nerve stimulation for faecal incontinence in the UKBritish Journal of Surgery, 2004
- Fecal incontinence quality of life scaleDiseases of the Colon & Rectum, 2000
- Electrical stimulation of sacral spinal nerves for treatment of faecal incontinenceThe Lancet, 1995
- Etiology and management of fecal incontinenceDiseases of the Colon & Rectum, 1993