A phantom based evaluation of the dose prediction and effects in treatment plans, when calculating on a direct density CT reconstruction
Open Access
- 16 March 2020
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 21 (3) , 52-61
Abstract
In radiation therapy, a Computed Tomography (CT) image is needed for an accurate dose calculation. To allow such a calculation, the CT image values have to be converted into relative electron densities. Thus, standard procedure is to calibrate the CT numbers to relative electron density (RED) by using a phantom with known composition inserts. This calibration curve is energy and CT dependent, therefore most radiotherapy CT acquisitions are obtained with 120 kVp, as each tube voltage needs an additional calibration curve. The commercially available DirectDensityTM (DD) reconstruction algorithm presents a reconstruction implementation without any dependence on the tube voltage. In comparison, it allows a calibration curve that is directly proportional to the RED, reducing the need of more than one calibration curve. This could potentially optimize CT acquisitions and reducing the dose given to the patient. Three different phantoms were used to evaluate the DirectDensityTM algorithm in simple and anthropomorphic geometries, as well as setups with metal implants. Scans with the DD algorithm were performed for 80, 100, 120, and 140 kVp. As reference a scan with the standard 120 kVp scan was used. Radiotherapy photon plans were optimized and calculated on the reference image and then transferred to the DD images, where they were recalculated. The dose distributions obtained this way were compared to the reference dose. Differences were found mainly in pure air and high density materials such as bones. The difference of the mean dose was below 0.7%, in most cases below 0.4%. No indication was found that the algorithm is corrupted by metal inserts, enabling the application for all clinical cases. This algorithm offers more variability in CT parameters for radiation therapy and thus a more personalized image acquisition with a high image quality and a lower dose exposure at a robust clinical workflow.Keywords
This publication has 10 references indexed in Scilit:
- Improving CT quality with optimized image parameters for radiation treatment planning and delivery guidancePhysics and Imaging in Radiation Oncology, 2017
- Clinical evaluation of a novel CT image reconstruction algorithm for direct dose calculationsPhysics and Imaging in Radiation Oncology, 2017
- Dosimetric Evaluation of Direct Electron Density Computed Tomography Images for Simplification of Treatment Planning WorkflowInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2016
- CT of the Abdomen with Reduced Tube Voltage in Adults: A Practical ApproachRadioGraphics, 2015
- Dosimetric comparison of Acuros XB, AAA, and XVMC in stereotactic body radiotherapy for lung cancerMedical Physics, 2014
- Experimental validation of the Eclipse AAA algorithmJournal of Applied Clinical Medical Physics, 2007
- Dose reduction in CT by on-line tube current control: principles and validation on phantoms and cadaversEuropean Radiology, 1999
- A technique for the quantitative evaluation of dose distributionsMedical Physics, 1998
- The calibration of CT Hounsfield units for radiotherapy treatment planningPhysics in Medicine & Biology, 1996
- An electron density calibration phantom for CT-based treatment planning computersMedical Physics, 1992