EDTA ‐induced pseudothrombocytosis and citrate‐induced platelet agglutination in a patient with Waldenstrom macroglobulinemia
Open Access
- 20 June 2017
- journal article
- research article
- Published by Wiley in Clinical Case Reports
- Vol. 5 (8) , 1243-1247
Abstract
Key Clinical Message: Hematology analyzers sometimes generate spurious results. A patient had EDTA ‐induced pseudothrombocytosis and platelet agglutination in citrate blood samples. This case verifies that addition of 1% paraformaldehyde to the citrate tubes can prevent platelet clumping. Further, it illustrates the advantages of having access to more than one platelet count method.Keywords
This publication has 27 references indexed in Scilit:
- Spurious counts and spurious results on haematology analysers: a review. Part I: plateletsInternational Journal of Laboratory Hematology, 2007
- The most accurate platelet count on the Sysmex XE-2100. Optical or impedance?Clinical and Laboratory Haematology, 2004
- Platelet Counting by the Coulter LH 750, Sysmex XE 2100, and Advia 120American Journal of Clinical Pathology, 2002
- Laboratory Identification of Cryoglobulinemia From Automated Blood Cell Counts, Fresh Blood Samples, and Blood FilmsAmerican Journal of Clinical Pathology, 2002
- Reduction of the Platelet Review Rate Using the Two-Dimensional Platelet MethodAmerican Journal of Clinical Pathology, 2001
- Improved Platelet Counting Using Two-Dimensional Laser Light ScatterAmerican Journal of Clinical Pathology, 2000
- Automated counting of platelets on the Bayer ADVIATTMM 120 analyserClinical and Laboratory Haematology, 1999
- Coulter S Plus STKR histograms detect spurious elevation of leucocyte and platelet counts associated with cryoglobulinaemiaAnnals of Hematology, 1989
- EDTA-dependent pseudothrombocytopaenia: A 12-month epidemiological studyScandinavian Journal of Clinical and Laboratory Investigation, 1988
- Pseudothrombocytopenia: A Laboratory Artifact With Potentially Serious ConsequencesMayo Clinic Proceedings, 1984