Characterization and Comparison of the 10-2 SITA-Standard and Fast Algorithms
Open Access
- 1 January 2012
- journal article
- clinical trial
- Published by Wiley in The Scientific World Journal
- Vol. 2012, 1-4
Abstract
Purpose: To compare the 10-2 SITA-standard and SITA-fast visual field programs in patients with glaucoma. Methods: We enrolled 26 patients with open angle glaucoma with involvement of at least one paracentral location on 24-2 SITA-standard field test. Each subject performed 10-2 SITA-standard and SITA-fast tests. Within 2 months this sequence of tests was repeated. Results: SITA-fast was 30% shorter than SITA-standard (5.5±1.1 vs 7.9±1.1 minutes, 𝑃<0.001). Mean MD was statistically significantly higher for SITA-standard compared with SITA-fast at first visit (Δ=0.3dB, 𝑃=0.017) but not second visit. Inter-visit difference in MD or in number of depressed points was not significant for both programs. Bland-Altman analysis showed that clinically significant variations can exist in individual instances between the 2 programs and between repeat tests with the same program. Conclusions: The 10-2 SITA-fast algorithm is significantly shorter than SITA-standard. The two programs have similar long-term variability. Average same-visit between-program and same-program between-visit sensitivity results were similar for the study population, but clinically significant variability was observed for some individual test pairs. Group inter- and intra-program test results may be comparable, but in the management of the individual patient field change should be verified by repeat testing.Keywords
This publication has 14 references indexed in Scilit:
- Long-term Survival of Central Visual Field in End-Stage GlaucomaOphthalmology, 2008
- Effect of Eye Testing Order on Automated Perimetry Results Using the Swedish Interactive Threshold Algorithm Standard 24-2Archives of Ophthalmology (1950), 2006
- Comparison of Glaucomatous Visual Field Defects Using Standard Full Threshold and Swedish Interactive Threshold AlgorithmsArchives of Ophthalmology (1950), 2002
- Sensitivity of Swedish interactive threshold algorithm compared with standard full threshold algorithm in Humphrey visual field testingOphthalmology, 2000
- The SITA perimetric threshold algorithms in glaucoma.1999
- Comparing significance and magnitude of glaucomatous visual field defects using the SITA and Full Threshold strategiesActa Ophthalmologica Scandinavica, 1999
- Evaluation of a new threshold visual field strategy, SITA, in normal subjectsActa Ophthalmologica Scandinavica, 1998
- The learning and fatigue effect in automated perimetryGraefe's Archive for Clinical and Experimental Ophthalmology, 1991
- Use of a central 10 degrees field and size V stimulus to evaluate and monitor small central islands of vision in end stage glaucoma.British Journal of Ophthalmology, 1991
- Changes in differential threshold in patients with glaucoma during prolonged perimetry.British Journal of Ophthalmology, 1983