Interpolation Scan Conversion in Pulse-Echo Ultrasound
Authors:
- 1 October 1982
- journal article
- research article
- Published by SAGE Publications in Ultrasonic Imaging
- Vol. 4 (4) , 297-310
Abstract
Some disadvantages of current digital scan conversion techniques are reviewed. An improved system is described which uses linear interpolation between available sampled data points to provide an estimate of the echo reflectivity at each grid point in the image. This approach avoids the problems of “missed pixels” inherent in the current approach, and the need to provide some smoothing or blurring function to remove them. The interpolated scan conversion allows data line spacing to be increased, leading to a reduced scan time for static scans and increased frame-rate for real-time, while retaining all the resolution available from the ultrasound system.Keywords
This publication has 7 references indexed in Scilit:
- Measurement of velocity of propagation from ultrasonic pulse-echo dataUltrasound in Medicine & Biology, 1982
- An Investigation of Computerized Tomography by Direct Fourier Inversion and Optimum InterpolationIEEE Transactions on Biomedical Engineering, 1981
- Computer Reconstruction Techniques in Compound Scan Pulse-Echo ImagingUltrasonic Imaging, 1981
- Direct Fourier reconstruction in computer tomographyIEEE Transactions on Acoustics, Speech, and Signal Processing, 1981
- Digital scan converters in diagnostic ultrasound imagingProceedings of the IEEE, 1979
- The UI Octoson — A New Class of Ultrasonic EchoscopeAustralasian Radiology, 1977
- A digital signal processing approach to interpolationProceedings of the IEEE, 1973