Multi‐Frequency Electrocochleography Results in Fewer Drop Alarms During Cochlear Implant Insertion
Open Access
- 5 May 2026
- journal article
- research article
- Published by Wiley in The Laryngoscope
- Vol. 136 (10) , 4474-4481
Abstract
Objective: To evaluate intracochlear electrocochleography (ECochG) amplitude parameters during cochlear implantation (CI) using a novel multi‐frequency ECochG algorithm. Methods: A multi‐institutional, prospective cohort study was performed at 18 high‐volume CI centers. The inclusion criteria were adults with sensorineural hearing loss and audiometric thresholds of ≤ 90 dB hearing level at 500 Hz undergoing CI with Advanced Bionics (Valencia, CA) Ultra 3D devices between 2024 and 2025. ECochG recordings were performed with simultaneous multi‐frequency stimulation of four frequencies between 125 and 4000 Hz during cochlear implant insertion. Concurrent multi‐frequency recording allowed extraction of amplitude and phase of each frequency individually. Post hoc analysis was performed to determine the difference in the number of drop alarms between single‐ and multi‐frequency ECochG. An ECochG amplitude drop of 6 dB was defined as a drop alarm. Insertion track patterns were compared between single‐ and multi‐frequency ECochG. Results: One hundred ninety‐five ears were included. Mean number of drop alarms for the single‐frequency algorithm was 1.72 (95% CI: 1.52, 1.92; median 1) compared to 0.42 (95% CI: 0.31, 0.53; median 0) for multi‐frequency; p < 0.001. The number of Type C patterns (rise in amplitude during insertion followed by a drop) decreased with the multi‐frequency ECochG algorithm compared to the single‐frequency ECochG algorithm. The number of Type D patterns (no‐response) decreased, indicating that multi‐frequency ECochG generated more responses across the cochlea than single‐frequency ECochG. Conclusions: A novel multi‐frequency ECochG algorithm during CI is associated with fewer drop alarms and altered insertion track patterns, which may provide a more accurate assessment of the cochlear microenvironment. Level of Evidence: 3.Keywords
This publication has 23 references indexed in Scilit:
- Monitoring Cochlear Health With Intracochlear Electrocochleography During Cochlear Implantation: Findings From an International Clinical InvestigationEar & Hearing, 2022
- Comparison of speech recognition in cochlear implant patients with and without residual hearing: A review of indicationsCochlear Implants International, 2021
- Toward a Better Understanding of Electrocochleography: Analysis of Real-Time RecordingsEar & Hearing, 2020
- Residual Cochlear Function in Adults and Children Receiving Cochlear Implants: Correlations With Speech Perception OutcomesEar & Hearing, 2019
- Intracochlear Electrocochleography: Response Patterns During Cochlear Implantation and Hearing PreservationEar & Hearing, 2018
- Multicenter US Clinical Trial With an Electric-Acoustic Stimulation (EAS) System in Adults: Final OutcomesOtology & Neurotology, 2018
- Patterns Seen During Electrode Insertion Using Intracochlear Electrocochleography Obtained Directly Through a Cochlear ImplantOtology & Neurotology, 2017
- Assessment of Cochlear Trauma During Cochlear Implantation Using Electrocochleography and Cone Beam Computed TomographyOtology & Neurotology, 2016
- Cochlear Implantation With Hearing Preservation Yields Significant Benefit for Speech Recognition in Complex Listening EnvironmentsEar & Hearing, 2013
- World Medical Association Declaration of HelsinkiJAMA, 2000