Optimal types of probe, and tissue Doppler frame rates, for use during tissue Doppler recording and off-line analysis of strain and strain rate in neonates at term
Authors:
- 1 April 2008
- journal article
- research article
- Published by Cambridge University Press (CUP) in Cardiology in the Young
- Vol. 18 (5) , 502-511
Abstract
Measurements of strain and strain rate, obtained by tissue Doppler might provide new parameters for assessing cardiac function in neonates. The noise-to-signal ratio is high. We investigated the effect of the frequency of the probe used, and the settings for tissue Doppler frame rate, on the noise in the analyses in three series of tissue Doppler images. In the first series, we used the 5S probe, with a frequency of 2.4 MHz, and the default frame rate. We used the10S probe, with a frequency of 8.0 MHz, in the other two series, one with a low and one with the default frame rate. The noise was lower using the 5S rather than the 10S probe, and lower when using the low frame rate rather than the default rate with the 10S probe. Using the settings eligible for two segment analyses with the lowest noise for each series, the noise was from 36 to 42% higher when using the 10S probe at default frame rate, and from 13 to 14% higher when using the 10S probe at low frame rate compared to the 5S probe at default frame rate. There were no differences in peak systolic strain or strain rate between the series. We found, therefore, that use of the 5S probe with the default setting for frame rate, along with a length of 1 mm and width of 2 mm for the region of interest, and a strain length of 10 mm, provided the optimal settings for two-segment analyses in this study.Keywords
This publication has 11 references indexed in Scilit:
- Feasibility and reliability of strain and strain rate measurement in neonates by optimizing the analysis parameters settingsUltrasound in Medicine & Biology, 2007
- Recovery of Stunned Myocardium in Acute Myocardial Infarction Quantified by Strain Rate Imaging: A Clinical StudyJournal of the American Society of Echocardiography, 2005
- Noninvasive quantification of regional myocardial function using Doppler-derived velocity, displacement, strain rate, and strain in healthy volunteers: effects of agingJournal of the American Society of Echocardiography, 2004
- Quantitative Assessment of Intrinsic Regional Myocardial Deformation by Doppler Strain Rate Echocardiography in HumansCirculation, 2002
- Quantification of regional left and right ventricular radial and longitudinal function in healthy children using ultrasound-based Strain Rate and Strain ImagingJournal of the American Society of Echocardiography, 2002
- Strain rate imaging in normal and reduced diastolic function: Comparison with pulsed doppler tissue imaging of the mitral annulusJournal of the American Society of Echocardiography, 2001
- Strain Rate Imaging by Ultrasonography in the Diagnosis of Coronary Artery DiseaseJournal of the American Society of Echocardiography, 2000
- Regional Strain and Strain Rate Measurements by Cardiac Ultrasound: Principles, Implementation and LimitationsEuropean Heart Journal - Cardiovascular Imaging, 2000
- Assessment of Regional Longitudinal Myocardial Strain Rate Derived from Doppler Myocardial Imaging Indexes in Normal and Infarcted MyocardiumJournal of the American Society of Echocardiography, 2000
- Real-Time Strain Rate Imaging of the Left Ventricle by UltrasoundJournal of the American Society of Echocardiography, 1998