The VeinViewer Vascular Imaging System Worsens First-Attempt Cannulation Rate for Experienced Nurses in Infants and Children with Anticipated Difficult Intravenous Access
Authors:
- Peter SzmukPSPeter Szmuk
- Jeffrey SteinerJSJeffrey Steiner
- Radu B. PopRPRadu B. Pop
- Alan Farrow-GillespieAFAlan Farrow-Gillespie
- Edward J. MaschaEMEdward J. Mascha
- Daniel I. SesslerDSDaniel I. Sessler
- 1 May 2013
- journal article
- research article
- Published by Wolters Kluwer Health in Anesthesia & Analgesia
- Vol. 116 (5) , 1087-1092
Abstract
The VeinViewer (Luminetx, Memphis, TN) helps identify veins by projecting an image of subcutaneous vasculature on the skin surface. We tested the primary hypothesis that VeinViewer use improves cannulation success by skilled nurses in pediatric patients with anticipated difficult IV access. A secondary goal was to evaluate the relationship between obesity and cannulation success. Patients aged 0 to 18 years were included. Anticipated cannulation difficulty was evaluated with the difficult IV access score. All cannulations were performed by members of the Intravenous Access Team. Patients were randomized to: (1) routine IV catheter insertion; or (2) insertion facilitated by the VeinViewer. The primary outcome was first-attempt insertion success. The proportion of successful insertions was evaluated using Cochran-Mantel-Haenszel χ(2) analysis to adjust for any imbalanced baseline variables. The effect of obesity on cannulation success was evaluated with multivariable logistic regression. Two hundred ninety-nine patients (49%) were randomly assigned to VeinViewer and 301 (51%) to routine cannulation. First-attempt cannulation success was 47% in patients assigned to VeinViewer vs 62% in patients assigned to routine cannulation, with an adjusted relative "risk" (95% confidence interval), of 0.76 (0.63-0.91). The Z-statistic of -3.6 crossed the "harm" boundary (Z < -2.41), with corresponding P value of 0.0003. The trial was stopped on statistical grounds since the harm boundary for the primary outcome was crossed. There was no association between first-attempt success and the 4-level categorization of obesity after adjusted for baseline variables (P = 0.94). The VeinViewer worsened first-attempt IV insertion success by skilled nurses. Surprisingly, first-attempt success for IV cannulation was not worsened by obesity.Keywords
This publication has 18 references indexed in Scilit:
- VeinViewer‐assisted Intravenous Catheter Placement in a Pediatric Emergency DepartmentAcademic Emergency Medicine, 2011
- Efficacy of a Near-Infrared Light Device in Pediatric Intravenous CannulationPediatric Emergency Care, 2011
- Real-Time Ultrasound-Guided Peripheral Vascular Access in Pediatric PatientsAnesthesia & Analgesia, 2010
- Transillumination by light‐emitting diode facilitates peripheral venous cannulations in infants and small childrenActa Anaesthesiologica Scandinavica, 2010
- Comparing peripheral venous access between obese and normal weight childrenPediatric Anesthesia, 2010
- Balance diagnostics for comparing the distribution of baseline covariates between treatment groups in propensity‐score matched samplesStatistics in Medicine, 2009
- A Vein Entry Indicator Device for Facilitating Peripheral Intravenous Cannulation in Children: A Prospective, Randomized, Controlled TrialAnesthesia & Analgesia, 2008
- Derivation of the DIVA ScorePediatric Emergency Care, 2008
- Introduction of a “Safety” Intravenous Catheter for Use in an Emergency Department: A Pediatric Hospital's ExperienceInfection Control & Hospital Epidemiology, 1992
- Group sequential designs using a family of type I error probability spending functionsStatistics in Medicine, 1990