The Importance of Off-Axis Loading and Articulation in Loosening of Cemented Glenoids
Open Access
- 1 August 2026
- journal article
- research article
- Published by MDPI AG in Biomechanics
- Vol. 6 (3) , 70
Abstract
Background: Glenoid loosening is the primary cause of failure in anatomic Total Shoulder Arthroplasty (aTSA). Current ASTM standards evaluate loosening via standard orthogonal vectors using non-continuous monitoring, which offer limited insight into the temporal evolution of loosening along physiological, off-axis wear vectors. This study demonstrates the feasibility and utility of a novel video-based system for continuous monitoring of glenoid loosening under clinically relevant, off-axis wear vectors and subluxation translation (ST) distances. Methods: A custom electromechanical apparatus cyclically displaced a humeral head against 18 cemented 5-peg glenoids over 150,000 cycles across three orientations: superior-inferior (SI), anterior–posterior (AP), and 45-degree angle (45°), and two ST distances (75% and 105% of maximum translation). Each condition had n = 3 glenoids. A video-based motion tracking system continuously monitored micromotion and relative motion at the bone-implant interface against a 150 µm loosening threshold. Differences were analyzed using Kruskal–Wallis tests and non-parametric effect sizes (η2). Results: The system successfully captured non-linear loosening behaviors. Final relative motion at 105% ST was SI (2476.0 ± 1882.3 µm), 45° (239.6 ± 51.9 µm), and AP (265.2 ± 99.7 µm), with all groups exceeding the 150 µm threshold. At 75% ST, relative motion was SI (175.7 ± 140.7 µm), 45° (24.3 ± 69.3 µm), and AP (33.1 ± 18.4 µm). While Kruskal–Wallis tests showed non-significant differences across orientations due to small sample sizes (p = 0.11), large effect sizes (η2 = 0.393) indicated practically meaningful differences, with SI showing the greatest motion. Conclusions: This study demonstrated the feasibility of a novel method of using a video-based system for the continuous monitoring of glenoid loosening under varying mechanical conditions, provides a comprehensive platform for evaluating the biomechanics of implant loosening, and suggests that testing along intermediate axes and at higher subluxation levels may be necessary to accurately predict in vivo performance.Keywords
This publication has 35 references indexed in Scilit:
- Cadaveric study validating in vitro monitoring techniques to measure the failure mechanism of glenoid implants against clinical CTJournal of Orthopaedic Research, 2018
- Stability of two versus three peripheral pegs of the glenoid component in modern total shoulder arthroplastyInternational Orthopaedics, 2017
- Posterior Glenoid Wear in Total Shoulder Arthroplasty: Eccentric Anterior Reaming Is Superior to Posterior AugmentClinical Orthopaedics and Related Research, 2015
- A biomechanical analysis of initial fixation options for porous-tantalum-backed glenoid componentsJournal of Shoulder and Elbow Surgery, 2013
- Failure mechanism of the all-polyethylene glenoid implantJournal of Biomechanics, 2010
- The glenoid in shoulder arthroplastyJournal of Shoulder and Elbow Surgery, 2009
- Static posterior humeral head subluxation and total shoulder arthroplastyJournal of Shoulder and Elbow Surgery, 2009
- Glenoid loosening after total shoulder arthroplasty: An in vitro CT‐scan studyJournal of Orthopaedic Research, 2009
- Cross-linked glenoid prosthesis: A wear comparison to conventional glenoid prosthesis with wear particulate analysisJournal of Shoulder and Elbow Surgery, 2009
- Total shoulder arthroplasty: Long-term survivorship, functional outcome, and quality of lifeJournal of Shoulder and Elbow Surgery, 2005