Mid-Term Comparative Follow-Up After Aortic Valve Replacement with Carpentier-Edwards and Pericarbon Pericardial Prostheses

Authors:
  • CSChristine Savoye
    Christine Savoye
  • EMEugene P. McFadden
    Eugene P. McFadden
  • DGDaniel Grandmougin
    Daniel Grandmougin
  • HCHubert-François Carton
    Hubert-François Carton
  • JHJean-Luc Hennequin
    Jean-Luc Hennequin
  • ADArnaud Dubar
    Arnaud Dubar
  • GFGeorges Fayad
    Georges Fayad
  • HWHenri Warembourg
    Henri Warembourg

Abstract
Background—The first generation of pericardial valves had a high rate of premature deterioration. The aim of this study was to compare the outcome after aortic valve replacement with second generation pericardial prostheses (Pericarbon and Carpentier-Edwards). Methods and Results—Between 1987 and 1994, 162 patients underwent aortic valve replacement with either a Pericarbon (n=81, 69±11 years) or a Carpentier-Edwards (n=81, 70±11 years) pericardial prosthesis. Mean follow-up was 4.4±2.7 years for Pericarbon and 4.8±2.4 years for Carpentier-Edwards valves (P=0.27), giving a total follow-up of 745 patient-years. Thirty-day mortality and 5-year actuarial survival were, respectively, 6.2% and 63.2±5.7% in the Pericarbon group and 6.2% and 63.5±5.6% in the Carpentier-Edwards group. At 8 years, freedom from (and linearized rates per patient-year) thromboembolism, structural failure, and all valve-related events were, respectively, 91.8±3.6% (1.4%), 76.9±8.7% (2.5%), and 58.4±9.3% (5.6%) in the Pericarbon group and 94.4±2.7% (1%), 100% (0%, PPConclusions—Eight years after aortic valve replacement, Pericarbon pericardial prostheses compared unfavorably with Carpentier-Edwards pericardial prostheses, with a high incidence of structural valve failure and reoperation.

This publication has 7 references indexed in Scilit: