The discrepancy between human peripheral nerve chronaxie times as measured using magnetic and electric field stimuli: the relevance to MRI gradient coil safety
Authors:
- Timothy J SchollTSTimothy J Scholl
- Blaine A ChronikBCBlaine A Chronik
- 17 September 2009
- journal article
- clinical trial
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 54 (19) , 5965-5979
Abstract
Peripheral nerve stimulation (PNS) resulting from electric fields induced from the rapidly changing magnetic fields of gradient coils is a concern in MRI. Nerves exposed to either electric fields or changing magnetic fields would be expected to display consistent threshold characteristics, motivating the direct application of electric field exposure criteria from the literature to guide the development of gradient magnetic field exposure criteria for MRI. The consistency of electric and magnetic field exposures was tested by comparing chronaxie times for electric and magnetic PNS curves for 22 healthy human subjects. Electric and magnetic stimulation thresholds were measured for exposure of the forearm using both surface electrodes and a figure-eight magnetic coil, respectively. The average chronaxie times for the electric and magnetic field conditions were 109 +/- 11 micros and 651 +/- 53 micros (+/-SE), respectively. We do not propose that these results call into question the basic mechanism, namely that rapidly switched gradient magnetic fields induce electric fields in human tissues, resulting in PNS. However, this result does motivate us to suggest that special care must be taken when using electric field exposure data from the literature to set gradient coil PNS safety standards in MRI.Keywords
This publication has 27 references indexed in Scilit:
- Accuracy Limitations of Chronaxie ValuesIEEE Transactions on Biomedical Engineering, 2004
- Simple anatomical measurements do not correlate significantly to individual peripheral nerve stimulation thresholds as measured in MRI gradient coilsJournal of Magnetic Resonance Imaging, 2003
- Comparison of the threshold for peripheral nerve stimulation during gradient switching in whole body MR systemsJournal of Magnetic Resonance Imaging, 2002
- A comparison between human magnetostimulation thresholds in whole‐body and head/neck gradient coilsMagnetic Resonance in Medicine, 2001
- Simple linear formulation for magnetostimulation specific to MRI gradient coilsMagnetic Resonance in Medicine, 2001
- Current FDA Guidance for MR Patient Exposure and Considerations for the FutureaAnnals of the New York Academy of Sciences, 1992
- Physiological Effects of Fast Oscillating Magnetic Field GradientsJournal of Computer Assisted Tomography, 1991
- Magnetic Stimulation of the Peripheral Nervous SystemJournal Of Clinical Neurophysiology, 1991
- Sensory stimulation by time‐varying magnetic fieldsMagnetic Resonance in Medicine, 1990
- Focal stimulation of human peripheral nerve with the magnetic coil: A comparison with electrical stimulationExperimental Neurology, 1989