Fundamental study of ion trapping and multiplexing using drift tube-ion mobility time-of-flight mass spectrometry for non-targeted metabolomics
- 13 July 2019
- journal article
- research article
- Published by Springer Nature in Analytical and Bioanalytical Chemistry
- Vol. 411 (24) , 6265-6274
Abstract
This study of ion accumulation/release behavior relevant to ion mobility–mass spectrometry (IM-MS) as employed for non-targeted metabolomics involves insight from theoretical studies, and controlled reference experiments involving measurement of low and high molecular mass metabolites in varying concentrations within a complex matrix (yeast extracts). Instrumental settings influencing ion trapping (accumulation time) and release conditions in standard and multiplexed operation have been examined, and translation of these insights to liquid chromatography (LC) in combination with drift tube IM-MS measurements has been made. The focus of the application is non-targeted metabolomics using carefully selected samples to allow quantitative interpretations to be made. Experimental investigation of the IM-MS ion utilization efficiency particularly focusing on the use of the Hadamard transform multiplexing with 4-bit pseudo-random pulsing sequence for assessment of low and high molecular mass metabolites is compared with theoretical modeling of gas-phase behavior of small and large molecules in the IM trapping funnel. Increasing the trapping time for small metabolites with standard IM-MS operation is demonstrated to have a deleterious effect on maintaining a quantitative representation of the metabolite abundance. The application of these insights to real-world non-targeted metabolomics assessment of intracellular extracts from biotechnologically relevant production processes is presented, and the results were compared to LC×IM-MS measurements of the same samples. Spiking of a uniformly 13C-labeled yeast extract (as a standard matrix) with varying amounts of natural metabolites is used to assess the linearity and sensitivity according to the instrument mode of operation (i.e., LC-MS, LC×IM-MS, and LC×[multiplexed]IM-MS). When comparing metabolite quantification using standard and multiplexed operation, sensitivity gain factors of 2–8 were obtained for metabolites with m/z below 250. Taken together, the simulation and experimental results of this study provide insight for optimizing measurement conditions for metabolomics and highlight the need for implementation of multiplexing strategies using short trapping times as relative quantification (e.g., in the context with non-targeted differential analysis) with sufficient sensitivity and working range is a requirement in this field of application.Keywords
This publication has 21 references indexed in Scilit:
- Ion mobility-derived collision cross section database: Application to mycotoxin analysisAnalytica Chimica Acta, 2018
- The potential of ion mobility–mass spectrometry for non-targeted metabolomicsCurrent Opinion in Chemical Biology, 2018
- Characterizing the lipid and metabolite changes associated with placental function and pregnancy complications using ion mobility spectrometry-mass spectrometry and mass spectrometry imagingPlacenta, 2017
- A structural examination and collision cross section database for over 500 metabolites and xenobiotics using drift tube ion mobility spectrometryChemical Science, 2017
- An Interlaboratory Evaluation of Drift Tube Ion Mobility–Mass Spectrometry Collision Cross Section MeasurementsAnalytical Chemistry, 2017
- Performance of a High-Pressure Liquid Chromatography-Ion Mobility-Mass Spectrometry System for Metabolic ProfilingAnalytical Chemistry, 2017
- Metabolomics and lipidomics using traveling-wave ion mobility mass spectrometryNature Protocols, 2017
- Comparing identified and statistically significant lipids and polar metabolites in 15‐year old serum and dried blood spot samples for longitudinal studiesRapid Communications in Mass Spectrometry, 2016
- Targeting the untargeted in molecular phenomics with structurally-selective ion mobility-mass spectrometryCurrent Opinion in Biotechnology, 2016
- Ion Mobility Derived Collision Cross Sections to Support Metabolomics ApplicationsAnalytical Chemistry, 2014