Optimum radiation source for radiation therapy of skin cancer
Authors:
Open Access
- 8 September 2015
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 16 (5) , 219-227
Abstract
Several different applicators have been designed for treatment of skin cancers, such as scalp, hand, and legs using Ir‐192 HDR brachytherapy sources (IR‐HDRS), miniature electronic brachytherapy sources (eBT), and external electron beam radiation therapy (EEBRT). Although, all of these methodologies may deliver the desired radiation dose to the skin, but the dose to the underlying bone may become the limiting factor for selection of the optimum treatment technique. In this project, dose to the underlying bone has been evaluated as a function of the radiation type, thickness of the bone, and thickness of the soft tissue on top of bone, assuming the same radiation dose delivery to the skin. These evaluations are performed using Monte Carlo (MC) simulation technique with MCNP5 code. The results of these investigations indicate that, for delivery of the same skin dose with a 50 keV eBT, 4 MeV or 6 MeV EEBRT techniques, the average doses received by the underlying bones are 5.31, 2, or 1.75 times the dose received from IR‐HDRS technique, respectively. These investigations indicate that, for the treatment of skin cancer condition with bone immediately beneath skin, the eBT technique may not be the most suitable technique, as it may lead to excessive bone dose relative to IR‐HDRS and 6 MeV or 4 MeV electron beams.PACS number: 87.53.Jw, 87.55.K‐Keywords
This publication has 20 references indexed in Scilit:
- A novel approach for superficial intraoperative radiotherapy (IORT) using a 50 kV X‐ray source: a technical and case reportJournal of Applied Clinical Medical Physics, 2014
- Complete response of endemic Kaposi sarcoma lesions with high-dose-rate brachytherapy: Treatment method, results, and toxicity using skin surface applicatorsBrachytherapy, 2013
- Nonmelanoma skin cancer treated with electronic brachytherapy: Results at 1 yearBrachytherapy, 2013
- Surface dose reduction from bone interface in kilovoltage X‐ray radiation therapy: a Monte Carlo study of photon spectraJournal of Applied Clinical Medical Physics, 2012
- New Technology in High-Dose-Rate Brachytherapy with Surface Applicators for Non-Melanoma Skin Cancer Treatment: Electronic Miniature X-Ray Brachytherapy2011
- Surface applicator calibration and commissioning of an electronic brachytherapy system for nonmelanoma skin cancer treatmenta)Medical Physics, 2010
- Design and evaluation of a HDR skin applicator with flattening filterMedical Physics, 2008
- Dosimetric characteristics of the Leipzig surface applicators used in the high dose rate brachy radiotherapyMedical Physics, 2004
- Results of radiotherapy for epithelial skin cancer of the pinna: the princess margaret hospital experience, 1982–1993International Journal of Radiation Oncology - Biology - Physics (IJROBP), 2000
- External irradiation of epithelial skin cancerInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1990