B0 and B1 inhomogeneities in the liver at 1.5 T and 3.0 T
- 26 October 2020
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 85 (4) , 2212-2220
Abstract
Purpose: The purpose of this work is to characterize the magnitude and variability of B0 and B1 inhomogeneities in the liver in large cohorts of patients at both 1.5 T and 3.0 T.Methods: Volumetric B0 and B1 maps were acquired over the liver of patients presenting for routine abdominal MRI. Regions of interest were drawn in the nine Couinaud segments of the liver, and the average value was recorded. Magnitude and variation of measured averages in each segment were reported across all patients.Results: A total of 316 B0 maps and 314 B1 maps, acquired at 1.5 T and 3.0 T on a variety of GE Healthcare MRI systems in 630 unique exams, were identified, analyzed, and, in the interest of reproducible research, de‐identified and made public. Measured B0 inhomogeneities ranged (5th‐95th percentiles) from −31.7 Hz to 164.0 Hz for 3.0 T (−14.5 Hz to 81.3 Hz at 1.5 T), while measured B1 inhomogeneities (ratio of actual over prescribed flip angle) ranged from 0.59 to 1.13 for 3.0 T (0.83 to 1.11 at 1.5 T).Conclusion: This study provides robust characterization of B0 and B1 inhomogeneities in the liver to guide the development of imaging applications and protocols. Field strength, bore diameter, and sex were determined to be statistically significant effects for both B0 and B1 uniformity. Typical clinical liver imaging at 3.0 T should expect B0 inhomogeneities ranging from approximately −100 Hz to 250 Hz (−50 Hz to 150 Hz at 1.5 T) and B1 inhomogeneities ranging from approximately 0.4 to 1.3 (0.7 to 1.2 at 1.5 T).Keywords
Funding Information
- Wisconsin Alumni Research Foundation
- GE Healthcare
- National Institutes of Health (K24‐DK102595, R01‐DK088925, R01‐DK100651, R01‐DK117354)
This publication has 38 references indexed in Scilit:
- R mapping in the presence of macroscopic B0 field variationsMagnetic Resonance in Medicine, 2011
- Quantitative assessment of liver fat with magnetic resonance imaging and spectroscopyJournal of Magnetic Resonance Imaging, 2011
- Addressing phase errors in fat‐water imaging using a mixed magnitude/complex fitting methodMagnetic Resonance in Medicine, 2011
- T1 independent, T2* corrected chemical shift based fat–water separation with multi‐peak fat spectral modeling is an accurate and precise measure of hepatic steatosisJournal of Magnetic Resonance Imaging, 2011
- Quantification of Hepatic Steatosis with T1-independent, T2*-corrected MR Imaging with Spectral Modeling of Fat: Blinded Comparison with MR SpectroscopyRadiology, 2011
- Magnetic Resonance Imaging Quantification of Liver IronMagnetic Resonance Imaging Clinics of North America, 2010
- B1 mapping by Bloch‐Siegert shiftMagnetic Resonance in Medicine, 2010
- Multiecho water‐fat separation and simultaneous R estimation with multifrequency fat spectrum modelingMagnetic Resonance in Medicine, 2008
- Joint estimation of water/fat images and field inhomogeneity mapMagnetic Resonance in Medicine, 2008
- Iterative decomposition of water and fat with echo asymmetry and least‐squares estimation (IDEAL): Application with fast spin‐echo imagingMagnetic Resonance in Medicine, 2005