Wavelet-based noise reduction for improved deconvolution of time-series data in dynamic susceptibility-contrast MRI
Authors:
- 10 May 2005
- journal article
- clinical trial
- Published by Springer Nature in Magnetic Resonance Materials in Physics, Biology and Medicine
- Vol. 18 (3) , 113-118
Abstract
Dynamic susceptibility-contrast (DSC) MRI requires deconvolution to retrieve the tissue residue function R(t) and the cerebral blood flow (CBF). In this study, deconvolution of time-series data was performed by wavelet-transform-based denoising combined with the Fourier transform (FT). Traditional FT-based deconvolution of noisy data requires frequency-domain filtering, often leading to excessive smoothing of the recovered signal. In the present approach, only a low degree of regularisation was employed while the major noise reduction was accomplished by wavelet transformation of data and Wiener-like filtering in the wavelet space. After inverse wavelet transform, the estimate of CBF·R(t) was obtained. DSC-MRI signal-versus-time curves (signal-to-noise ratios 40 and 100) were simulated, corresponding to CBF values in the range 10–60 ml/(min 100 g). Three shapes of the tissue residue function were investigated. The technique was also applied to six volunteers. Simulations showed CBF estimates with acceptable accuracy and precision, as well as independence of any time shift between the arterial input function and the tissue concentration curve. The grey-matter to white-matter CBF ratio in volunteers was 2.4±0.2. The proposed wavelet/FT deconvolution is robust and can be implemented into existing perfusion software. CBF maps from healthy volunteers showed high quality.Keywords
This publication has 18 references indexed in Scilit:
- Removing the effect of SVD algorithmic artifacts present in quantitative MR perfusion studiesMagnetic Resonance in Medicine, 2004
- ForWaRD: Fourier-Wavelet Regularized Deconvolution for Ill-Conditioned SystemsIEEE Transactions on Signal Processing, 2004
- Tracer arrival timing‐insensitive technique for estimating flow in MR perfusion‐weighted imaging using singular value decomposition with a block‐circulant deconvolution matrixMagnetic Resonance in Medicine, 2003
- Perfusion quantification using Gaussian process deconvolutionMagnetic Resonance in Medicine, 2002
- High resolution measurement of cerebral blood flow using intravascular tracer bolus passages. Part I: Mathematical approach and statistical analysisMagnetic Resonance in Medicine, 1996
- Quantification of regional cerebral blood flow and volume with dynamic susceptibility contrast-enhanced MR imaging.Radiology, 1994
- A deconvolution method for evaluating indicator-dilution curvesPhysics in Medicine & Biology, 1994
- Simultaneous MR Acquisition of Arterial and Brain Signal‐Time CurvesMagnetic Resonance in Medicine, 1992
- Perfusion imaging with NMR contrast agentsMagnetic Resonance in Medicine, 1990
- CEREBRAL BLOOD FLOW, BLOOD VOLUME AND OXYGEN UTILIZATIONBrain, 1990