Collapse of a Stent-Graft Following Treatment of a Traumatic Thoracic Aortic Rupture

Authors:

Abstract
Purpose: To report a collapsed stent-graft used to treat a traumatic aortic rupture. Case Report: A Gore TAG stent-graft was placed in a 20-year-old man with multiple injuries. Postimplantation computed tomographic angiography (CTA) demonstrated no contrast extravasation and total exclusion of the traumatic rupture. Routine CTA 3 months after implantation revealed a collapsed stent-graft located in the outer curve of the distal aortic arch. A Talent stent-graft was placed successfully within the collapsed prosthesis. Postimplantation CTA demonstrated no contrast extravasation and good apposition of the endograft to the aortic wall. At 6 months, the repair remains secure; there is no sign of graft collapse or endoleak. Conclusions: Collapse of stent-grafts can occur after treatment for traumatic aortic ruptures; endovascular methods can be used to restore a satisfactory luminal contour.