Validation of an experimental polyurethane model for biomechanical studies on implant supported prosthesis - tension tests
Open Access
- 1 June 2011
- journal article
- research article
- Published by FapUNIFESP (SciELO) in Journal of Applied Oral Science
- Vol. 19 (3) , 244-248
Abstract
The complexity and heterogeneity of human bone, as well as ethical issues, frequently hinder the development of clinical trials. The purpose of this in vitro study was to determine the modulus of elasticity of a polyurethane isotropic experimental model via tension tests, comparing the results to those reported in the literature for mandibular bone, in order to validate the use of such a model in lieu of mandibular bone in biomechanical studies. Forty-five polyurethane test specimens were divided into 3 groups of 15 specimens each, according to the ratio (A/B) of polyurethane reagents (PU-1: 1/0.5, PU-2: 1/1, PU-3: 1/1.5). Tension tests were performed in each experimental group and the modulus of elasticity values found were 192.98 MPa (SD=57.20) for PU-1, 347.90 MPa (SD=109.54) for PU-2 and 304.64 MPa (SD=25.48) for PU-3. The concentration of choice for building the experimental model was 1/1.Keywords
This publication has 26 references indexed in Scilit:
- Fracture Toughness and Fatigue Crack Propagation Rate of Short Fiber Reinforced Epoxy Composites for Analogue Cortical BoneJournal of Biomechanical Engineering, 2007
- Artificial composite bone as a model of human trabecular bone: The implant–bone interfaceJournal of Biomechanics, 2007
- An In Vitro Comparison of an Adjustable Bone Fixation SystemJournal of Oral and Maxillofacial Surgery, 2005
- Response of osteoblastic cells to titanium submitted to three different surface treatmentsBrazilian Oral Research, 2005
- Edentulation Alters Material Properties of Cortical Bone in the Human MandibleJournal of Dental Research, 2002
- Anisotropic elastic properties of cancellous bone from a human edentulous mandibleClinical Oral Implants Research, 2000
- Mechanotransduction and the functional response of bone to mechanical strainCalcified Tissue International, 1995
- Dental implants: materials and design considerations.1993
- Microbial differences in 2 clinically distinct types of failures of osseointegrated implantsClinical Oral Implants Research, 1991
- Basic biomechanics of dental implants in prosthetic dentistryThe Journal of Prosthetic Dentistry, 1989