Evidence to support magnetic resonance conditional labelling of all pacemaker and defibrillator leads in patients with cardiac implantable electronic devices
Open Access
- 26 August 2021
- journal article
- research article
- Published by Oxford University Press (OUP) in European Heart Journal
- Vol. 43 (26) , 2469-2478
Abstract
Aims: Many cardiac pacemakers and defibrillators are not approved by regulators for magnetic resonance imaging (MRI). Even following generator exchange to an approved magnetic resonance (MR)-conditional model, many systems remain classified ‘non-MR conditional’ due to the leads. This classification makes patient access to MRI challenging, but there is no evidence of increased clinical risk. We compared the effect of MRI on non-MR conditional and MR-conditional pacemaker and defibrillator leads. Methods and results: Patients undergoing clinical 1.5T MRI with pacemakers and defibrillators in three centres over 5 years were included. Magnetic resonance imaging protocols were similar for MR-conditional and non-MR conditional systems. Devices were interrogated pre- and immediately post-scan, and at follow-up, and adverse clinical events recorded. Lead parameter changes peri-scan were stratified by MR-conditional labelling. A total of 1148 MRI examinations were performed in 970 patients (54% non-MR conditional systems, 39% defibrillators, 15% pacing-dependent) with 2268 leads. There were no lead-related adverse clinical events, and no clinically significant immediate or late lead parameter changes following MRI in either MR-conditional or non-MR conditional leads. Small reductions in atrial and right ventricular sensed amplitudes and impedances were similar between groups, with no difference in the proportion of leads with parameter changes greater than pre-defined thresholds (7.1%, 95% confidence interval: 6.1–8.3). Conclusions: There was no increased risk of MRI in patients with non-MR conditional pacemaker or defibrillator leads when following recommended protocols. Standardizing MR conditions for all leads would significantly improve access to MRI by enabling patients to be scanned in non-specialist centres, with no discernible incremental risk.Keywords
Funding Information
- British Heart Foundation (FS/16/46/32187, HFHF_016)
- NIHR Biomedical Research Centres at University College London Hospital
- Barts Health NHS Trusts
This publication has 26 references indexed in Scilit:
- Provision of magnetic resonance imaging for patients with ‘MR-conditional’ cardiac implantable electronic devices: an unmet clinical needEP Europace, 2016
- Clinical value of cardiovascular magnetic resonance in patients with MR-conditional pacemakersEuropean Heart Journal - Cardiovascular Imaging, 2015
- Utilization and likelihood of radiologic diagnostic imaging in patients with implantable cardiac defibrillatorsJournal of Magnetic Resonance Imaging, 2015
- Full-Body MRI in Patients With an Implantable Cardioverter-DefibrillatorJACC, 2015
- Determining the Risks of Magnetic Resonance Imaging at 1.5 Tesla for Patients With Pacemakers and Implantable Cardioverter DefibrillatorsThe American Journal of Cardiology, 2012
- High Ventricular Lead Impedance of a DDD Pacemaker after Cranial Magnetic Resonance ImagingPacing and Clinical Electrophysiology, 2011
- Magnetic resonance imaging in patients with a pacemaker system designed for the magnetic resonance environmentHeart Rhythm, 2011
- In vivo heating of pacemaker leads during magnetic resonance imagingEuropean Heart Journal, 2004
- Modern Pacemaker and Implantable Cardioverter/Defibrillator Systems Can Be Magnetic Resonance Imaging SafeCirculation, 2004
- Effects of magnetic resonance imaging on cardiac pacemakers and electrodesAmerican Heart Journal, 1997