Study of SO2 measurement based on a dual optical path Fabry-Perot correlation spectroscopy

Authors:

Abstract
This paper investigates a method for measuring SO2 concentration using Fabry–Perot interferometer correlation spectroscopy. In this method, the experimental system is designed as a separated beam, with the beam entering the F–P cavity at two incidence angles simultaneously to match the peak and valley positions of the SO2 absorption cross-section. The system achieves a 2σ detection limit of 28.2 ppm·m(15 cm) at a sampling frequency of 10 Hz. An outfield comparison experiment with the differential optical absorption spectroscopy method shows good agreement for the simultaneous measurement of SO2 concentration from sulfur combustion, with a correlation coefficient of R2 = 0.93. This study introduces a non-dispersive, highly accurate, and fast gas detection technique.